background image

Glossary 

Operator Manual for the Baxter ExactaMix 2400 Compounder 

Page 202 

53007410  Rev. C 

Issues with the Weight and Load Cell 

On-screen Text 

Explanation 

Suggested Actions 

Expected Weight: 

<

calculated weight

> gm  

 

Actual Weight: <

actual 

weight

> gm  

 

Difference: <

weight 

difference

>%  

 

The final weight of this 

solution is outside of the 

acceptable limit of +/-5%  

 

Possible Cause: Unknown  

The final bag weight is out 

of range. 

 

After the compounder 

delivers all the 

ingredients, the weight of 

the patient bag differs 

from the expected weight 

by more than the 

acceptable difference. 

1.

 

Check that all the source containers are spiked 

properly. Refer to the steps for spiking a 

container, starting on Page 51. 

2.

 

Check that the pump rotor is clean. Refer to 

CLEANING THE COMPOUNDER on Page 101. 

3.

 

Check that the outlet tube is installed properly. 

Refer to Installing the New Valve Set on 

Page 45. 

4.

 

Calibrate the load cell. Refer to Calibrating the 

Load Cell on Page 37. 

5.

 

Calibrate the pump. Refer to Calibrating the 

Compounder on Page 69. 

6.

 

Compound a large-volume solution with at 

least 205 mL of water to make the compounder 

calibrate automatically. 

Expected Weight: 

<

calculated weight

> gm  

 

Actual Weight: <

actual 

weight

> gm  

 

Difference: <

weight 

difference

>%  

 

The final weight of this 

solution is within the 

acceptable limit of +/-5%, 

however some 

ingredients may not have 

delivered correctly.  

 

Possible Cause: 

<

ingredient name

> is 

possibly <

underweight / 

overweight

> by <

weight 

error

> grams  

An individual ingredient 

delivery is out of range. 

 

After the compounder 

delivers an ingredient, the 

weight of the patient bag 

differs from the expected 

weight by more than the 

acceptable difference. 

 

The compounder checks 

the weight after individual 

ingredient deliveries over 

100 mL.  

 

 

1.

 

Check that all the ingredients and inlets are 

correct. 

2.

 

Check that the tube set is installed properly. 

Refer to Installing the New Valve Set on 

Page 45 and Attaching the New Ingredients and 

Inlets on Page 48. 

3.

 

On the MixCheck Report, check for references 

to occlusions and bubbles. Refer to MixCheck 

Report on Page 161. Have a pharmacist check 

the accuracy. 

4.

 

Check that all the source containers are spiked 

properly. Refer to the steps for spiking a 

container, starting on Page 51. 

5.

 

Check that the pump rotor is clean. Refer to 

CLEANING THE COMPOUNDER on Page 101. 

6.

 

Calibrate the load cell. Refer to Calibrating the 

Load Cell on Page 37. 

7.

 

Calibrate the pump. Refer to Calibrating the 

Compounder on Page 69. 

8.

 

Compound a large-volume solution with at 

least 205 mL of water to make the compounder 

calibrate automatically. 

9.

 

Contact Baxter Technical Services to check that 

the flow factors are correct. Refer to Getting 

Help on Page 22. 

Bag out of range 

Dextrose is pumping too 

quickly or slowly. 

1.

 

Check that there are no environmental factors 

interfering with pumping dextrose. 

 

Keep the room temperature stable. 

 

Always allow refrigerated dextrose to return 

to room temperature before using it. 

2.

 

Check that the source container is spiked 

properly. Refer to the steps for spiking a 

container, starting on Page 51. 

3.

 

Check that the correct inlet is assigned to 

dextrose in the formulary. Refer to Adding or 

Editing a Product on Page 143. 

Summary of Contents for ExactaMix 2400 Compounder

Page 1: ...on order of a physician Fuse TERMS IN THIS MANUAL WARNING Indicates a risk of personal injury or patient harm if the instructions are not followed CAUTION Indicates a risk of damage to equipment or da...

Page 2: ......

Page 3: ...vigation 8 Menu Screen 8 Pump Screen 9 Tutorials 12 FEATURES 13 Summary of Features 13 Order Entry 13 Formulary 13 Ingredient Groups 14 Universal Ingredient 14 Configuration 15 Barcode Verification 15...

Page 4: ...Set 45 Preparing the New Ingredients 47 Attaching the New Ingredients and Inlets 48 Priming and Verifying 58 Verifying the Ingredient and Inlet Barcodes 59 Priming the Inlets and Verifying the Setup 6...

Page 5: ...and Errors 100 MAINTAINING THE COMPOUNDER 101 Calibrating the Load Cell 101 Changing the Tube Set 101 CLEANING THE COMPOUNDER 101 Calibrating the Compounder 103 Shutting Down and Starting Up the Compo...

Page 6: ...e Label 118 Setting Up the Security Options 119 General 119 Auto Logout 119 Password Expiration 120 Order Entry Serial Number 120 Barcode Reader 120 Default User 120 Cosignature 120 Solution Limit 121...

Page 7: ...Editing a Product 143 Deleting a Product 145 Viewing a Product s Usage Information 146 USING THE INGREDIENT GROUP EDITOR 147 Adding or Editing an Ingredient Group 148 Deleting an Ingredient Group 149...

Page 8: ...TING 193 Handling Interruptions and Errors 193 Issues with the Barcodes 193 Issues with the Formulas Ingredients and Configurations 197 Issues with the Calibration 201 Issues with the Weight and Load...

Page 9: ...the compounder withdraws a specified volume of each ingredient from its source container in a specified sequence and pumps each ingredient into a patient bag The finished solution is delivered to pat...

Page 10: ...3007410 Rev C COMPONENTS HARDWARE COMPONENTS The compounder consists of these main hardware components Hardware components NOTE The vial rack extension optional and the printer North America only are...

Page 11: ...s paused the valve actuators automatically close The occlusion detector detects occlusions blockages in the tube between the source containers and the detector The bubble detector detects air bubbles...

Page 12: ...d cell The display operates the compounder software and includes a touch screen for easy input The bottom of the display contains four USB ports which can be used to connect a barcode reader keyboard...

Page 13: ...holders and syringe holders attach to the vial rack Vial rack The base plate is the common base on which the compounder s components sit Base plate The laser printer used for printing reports and labe...

Page 14: ...refer to Inlets on Page 18 NOTE The valve set and inlets are collectively known as the tube set The destination bag is a sterile container that holds the fluid pumped from the source containers There...

Page 15: ...by Baxter Installing any software not approved by Baxter may change the operating parameters adversely affect the operation and create unsafe conditions Installing this software voids the compounder...

Page 16: ...8 53007410 Rev C Navigation On any screen or window that requires data entry tapping a field displays an on screen keyboard or number pad that allows you to enter characters On screen keyboard and nu...

Page 17: ...a to compound and manage ingredients Tools menu to set up options related to the system users security directories used for saving certain files and software maintenance Reports menu to view print and...

Page 18: ...the top of the screen Buttons appear on the left side of the screen Tapping Run starts the compounding process Tapping Menu displays the menu screen The total volume to be pumped for the order appears...

Page 19: ...a red X indicating that they need to be verified The ordered volume of the ingredient A number indicating the ingredient s place in the compounding sequence A vertical bar showing how much of the ingr...

Page 20: ...destination bag Horizontal marks across an inlet represent fluid indicating that this inlet has been primed Tutorials The compounder software includes step by step tutorials about setting up the comp...

Page 21: ...or and occlusion detector Can be immediately stopped by lifting the pump door Can track ingredient lot numbers and ingredient expiration dates Generates a MixCheck Report for each finished solution Ca...

Page 22: ...he common fluid pathway or within the patient bag in the absence of sufficient volume These ingredients are considered to be incompatible For example calcium and phosphate should not be mixed in their...

Page 23: ...tifies the port to which the inlet is attached These barcode labels are packaged with the valve set The compounder software can also generate a report that makes these labels available for printing Mo...

Page 24: ...s formula PRINTING OPTIONS The printer is used for printing reports and creating labels for inlets and source containers The printer can use standard 8 5 x 11 in 21 6 x 28 cm for reports and Avery lab...

Page 25: ...PAT FRM files send print jobs and back up the database Baxter does not support network related equipment nor activities related to setting up or troubleshooting network connectivity for the compounde...

Page 26: ...Large volume non vented containers that require a spike to vent air into the container 50 60 mL 25 30 mL Vented Micro Volume Inlet H938175 25 Small volume vials 5 6 mL 2 5 3 mL Micro Volume Inlet with...

Page 27: ...3000 mL H938741 50 ExactaMix Empty EVA Container 4000 mL H938142 50 ExactaMix Empty EVA Container 5000 mL H938143 50 ExactaMix Empty EVA Container dual chamber 1500 mL H938901 42 ExactaMix Empty EVA C...

Page 28: ...un EVA TPN Bag 1000 mL 5 pack EVA 1000 40 Lg Bore B Braun E1320 OV 5 B Braun EVA TPN Bag 2000 mL 5 pack EVA 2000 35 Lg Bore B Braun E1330 OV 5 B Braun EVA TPN Bag 3000 mL 5 pack EVA 3000 35 Lg Bore B...

Page 29: ...s have a large bore threaded fill port connector or a luer lock fill port connector VALVE SETS The following valve sets are available in North America Product Order Number Quantity Case Notes EM2400 V...

Page 30: ...rvices is available 24 hours day 7 days week at 1 800 678 2292 or COtechsupport baxter com Before you call for technical support 1 At the menu screen tap Help About The About window appears It provide...

Page 31: ...and inspect them to make sure that they are not damaged WARNING Do not use sharp objects to open the packaging Personal injury could result The compounder should be placed on a level and stable surfa...

Page 32: ...onnect or disconnect the display and load cell The power cord must be positioned so that the plug is easily accessible 6 Under the main module connect the following cord and cables Reach under the mai...

Page 33: ...t the main module rests fully on the base plate 8 Route the cables through the two routing holes pushing any excess length of the cables through the holes Routing a cable 9 Install the load cell a Pla...

Page 34: ...alling the display NOTE You can remove the display by pulling the locking pin out to the unlocked position then rotating the pin 90 degrees to keep it in this position while lifting the display 11 Ins...

Page 35: ...ble from the barcode reader to a USB port on the bottom of the display 13 If desired Connect the USB cable from the printer to a USB port on the display Connect an Ethernet cable to the Ethernet port...

Page 36: ...isplay forward or backward c Lock the lever by rotating it backward Lever on the right 15 Move the display to the desired position a Unlock the lever on the left by rotating it backward NOTE Pulling t...

Page 37: ...he bolt on the main module Installing the vial rack 19 Install the vial holders in the desired locations on the vial rack 20 Adjust the position of each top and bottom vial holder a Rotate the cam up...

Page 38: ...luminates Power light and power button 2 If the software does not start automatically double tap the Exacta Mix 2400 icon on the Windows desktop Tip Baxter recommends setting the software to start aut...

Page 39: ...age 120 When the software starts it performs self checks and briefly displays the following messages Other messages may also appear Do not cancel these operations Messages that appear at startup Next...

Page 40: ...ps before compounding LOGGING OUT When you have finished using the compounder or another user needs to log in you can log out of the software without shutting down the compounder At the menu screen ta...

Page 41: ...these options Shutdown the computer to shut down the software and turn off the compounder Reboot the computer to shut down the software turn off the compounder and restart the compounder Log on as a d...

Page 42: ...used only to turn the power on CAUTION You can reboot the display by pressing and holding the Reset button on the side of the display You can shut down the compounder main module and display by press...

Page 43: ...ways Tap Yes at the Confirm screen if it appears during startup Tap Compound Setup Wizard at the menu screen to access the Setup Wizard at any time NOTE The Compound menu also includes options that a...

Page 44: ...brating the load cell 2 Select Configuration allows you to select a configuration to use 3 Change Tube Set guides you through installing a new tube set and new ingredients for the selected configurati...

Page 45: ...n the calibration expires Using a 2 000 g weight that Baxter provides Tip Baxter recommends Using gloves to handle the weight to minimize the accumulation of oils and dust Storing the weight in its st...

Page 46: ...uch the load cell during the calibration the calibration will not be accurate 2 Make sure that There is no weight on the load cell There is nothing touching any part of the load cell for example there...

Page 47: ...eight message tap OK This message appears and then disappears Message When calibration is finished one of these results occurs If you accessed the calibration procedure from the Setup Wizard screen yo...

Page 48: ...want to use a configuration that is different from the last one that was used The software automatically selects the last configuration that was used and a check mark appears next to Select Configura...

Page 49: ...axter ExactaMix 2400 Compounder 53007410 Rev C Page 41 2 In the Name list select the desired configuration 3 Tap OK Select Configuration screen NOTE When you select a different configuration you may n...

Page 50: ...WARNING To maintain delivery accuracy the tube set must be replaced after it has delivered 150 L of fluid or been installed for 24 hours whichever comes first Check that the materials of the inlets va...

Page 51: ...d be changed Tube set statistics and recommendations 2 Tap one of these options Tap Tube Set will be changed then continue with Removing the Expired Tube Set and Expired Ingredients on Page 44 NOTE Se...

Page 52: ...tainer from the vial rack or hanger and turn the container right side up 3 Lower the container below the height of the valve set and allow the fluid in the inlet to flow back into the source container...

Page 53: ...o Getting Help on Page 22 2 Remove the valve set from the packaging NOTE Always use aseptic technique when installing the valve set IMPORTANT Check the expiration date on the valve set before installi...

Page 54: ...the outlet tube into channel 1 around the pump rotor and into channels 2 and 3 as shown Move the pump rotor counterclockwise only Do not pull or stretch the outlet tube Routing the outlet tube a Make...

Page 55: ...ommends using only sterile BD 60 mL Luer syringes If you plan to use a different type contact Baxter Technical Services Refer to Getting Help on Page 22 1 Gather all the new ingredients Tip Baxter rec...

Page 56: ...nds working from left to right in the sequence of the port numbers 1 2 3 and so on You might find it helpful to remember the main steps of this process covered in more detail on the upcoming pages by...

Page 57: ...10 Rev C Page 49 Hang Source Containers screen no ingredients attached The ingredient detail window appears IMPORTANT Always view the ingredient detail window It includes details not visible on the in...

Page 58: ...ts of the complete part number For a list of complete part numbers refer to Inlets on Page 18 The complete part number also appears on the packaging materials for the inlet d Check that the Remainder...

Page 59: ...volumes resulting in patient harm After attaching the inlets visually check that the tubes are not kinked or plugged c Check that the inlet is not kinked or plugged d On the valve set locate the port...

Page 60: ...rt facing down NOTE This step helps to reduce the possibility of air bubbles entering the inlet tube Turning WARNING Failure to insert the spike completely into the bag port may restrict flow and caus...

Page 61: ...E This step helps to reduce the possibility of air bubbles entering the inlet tube Turning b Locate the shoulder of the spike Shoulder c Insert the spike fully into the bottle up to the shoulder of th...

Page 62: ...es entering the inlet tube Turning b Insert the spike fully into the vial Inserting c Push the bottom of the vial now facing up against the top holder Make sure that the vent faces you NOTE If the ven...

Page 63: ...inge on the vial rack by snapping the syringe flanges into the syringe holder Hanging 6 Label the inlet with the appropriate numbered inlet barcode label that was packaged with the valve set Attach th...

Page 64: ...ExactaMix 2400 Compounder Page 56 53007410 Rev C At the Hang Source Containers screen the color of the ingredient button becomes teal to indicate that the ingredient is attached and waiting to be prim...

Page 65: ...steps for all the ingredients you want to attach When all the ingredient buttons are teal you are finished changing the tube set 10 Tap OK Hang Source Containers screen all ingredients attached If you...

Page 66: ...up 1 At the Setup Wizard screen tap Prime and Verify Setup Wizard screen priming and verifying NOTE To prime and verify without using the Setup Wizard you can tap Compound Prime and Verify at the menu...

Page 67: ...efer to Attaching the New Ingredients and Inlets on Page 48 IMPORTANT This procedure requires barcode scanning to be enabled To enable barcode scanning and if desired to require it for verification re...

Page 68: ...p Baxter strongly recommends scanning from left to right or from right to left to prevent skipping any ports 1 Scan the barcode label on an inlet 2 Scan the barcode label on the corresponding source c...

Page 69: ...ct is not the specific product identified in the configuration but is the same ingredient a Warning message appears Tap Yes if you want to use the scanned product Message 3 Repeat the previous steps u...

Page 70: ...et to the correct port Incorrect setup could result in patient harm IMPORTANT The cosignature option requires Verification permissions For more information about user groups and permissions refer to S...

Page 71: ...f these ingredients can cause a precipitate which may block the common fluid pathway and require a replacement of the tube set If the configuration includes a lipid you should prime the Universal Ingr...

Page 72: ...tion in the title bar of the window is correct b Check that the Port Number is correct c Check that the Ingredient description matches the source container to be used d Check that the Remainder mL mat...

Page 73: ...heck that the product attached to the inlet matches the information on the screen c At the ingredient detail window tap Prime d Check that a calibration bag is attached e At the Priming pump message t...

Page 74: ...k that the inlet is attached to the correct port and prime it again After the inlet has been primed the ingredient detail window includes a Verify button and the Prime button becomes a Re Prime button...

Page 75: ...e ingredient has been primed On the ingredient button the red X becomes a green check mark indicating that the ingredient has been verified PRIME AND VERIFY screen one inlet primed and verified 11 Rep...

Page 76: ...Fluid path will be flushed with UI message tap OK Message The compounder flushes the common fluid pathway with the Universal Ingredient and tests for proper function of the occlusion detector If the t...

Page 77: ...the Setup Wizard screen a check mark now appears next to Prime and Verify at the Setup Wizard screen CALIBRATING THE COMPOUNDER This procedure calibrates the compounder s pump to ensure that it deliv...

Page 78: ...rrently on the load cell does not appear to be empty on Page 203 If the bag contains fluid continue with the next step 5 Check whether the bag has space for an additional 200 mL of fluid If the bag do...

Page 79: ...he type of bag you are using b Tap OK Select Bag Type window sample North American version IMPORTANT If any items touch the load cell during the calibration the calibration will not be accurate 7 Make...

Page 80: ...ion is something other than water the compounder automatically flushes the common fluid pathway with the identified UI This message appears and disappears Message 8 At the Pump calibration completed s...

Page 81: ...ing the tube 2 Attach the bag to the load cell Place the holes in the corners of the bag over the guide pins on the load cell 3 Route the bag s fill port through the load cell s fill port holder 4 Mak...

Page 82: ...ation Report button becomes active at the Setup Wizard screen At the Setup Wizard screen tap Authorization Report Setup Wizard screen viewing the Authorization Report NOTE You can view the Authorizati...

Page 83: ...by Scanning a Barcode Most facilities use this method IMPORTANT This method requires Order entry software on a separate computer This software must be able to produce both a PAT FRM file and a corres...

Page 84: ...C WARNING The code number for each product in the formula must exactly match the code number for that product in the compounder s formulary If a code number is assigned to one product in the order ent...

Page 85: ...with a unique ID but it does not create a 2D Formula Barcode PAT FRM file or a corresponding label with a barcode WARNING Formulas entered directly into the compounder should be checked by a pharmacis...

Page 86: ...nder Page 78 53007410 Rev C The pump screen appears The text DIRECT ENTRY appears on the left side Pump screen during direct entry 2 Obtain the formula from the pharmacist Refer to your facility s pro...

Page 87: ...Check its description concentration and so on b Enter the Ordered Volume of the ingredient to be delivered c Tap Save or tap Next to view the next ingredient detail window The Run button becomes avai...

Page 88: ...g Automatic Unloading of Formulas on Page 84 1 Attach the patient bag Refer to Attaching the Patient Bag on Page 85 2 At the pump screen tap Run 3 At the Formula Information window a Enter a Formula N...

Page 89: ...Compounder Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 81 Saving a Direct Entry Formula to Use Later 1 At the pump screen tap Menu 2 At the Information message tap OK...

Page 90: ...7410 Rev C 3 At the Formula Information window a Enter a Formula Name b If desired enter a new Serial Number Tip Baxter recommends not changing the serial number c Tap OK Formula Information window 4...

Page 91: ...lect Formula window 2 At the Select Formula window tap one of these filter options Show All Formulas to view all the formulas that are stored Show Unpumped Formulas to view all the formulas that have...

Page 92: ...he only exception is that the software does not unload the formula when you perform an auto addition Message The software usually unloads the formula when compounding is finished regardless of the out...

Page 93: ...ter recommends connecting the bag to the outlet tube before attaching it to the load cell to prevent twisting or straining the tube 2 Attach the bag to the load cell Place the holes in the corners of...

Page 94: ...ge 86 53007410 Rev C Compounding the Solution IMPORTANT This function requires Compounder permissions For more information about user groups and permissions refer to Setting Up the Users on Page 123 1...

Page 95: ...e 73 d Attach an empty bag Refer to Attaching the Patient Bag on Page 85 e Return to Step 1 Message WARNING If you tap Yes the compounder will reset the measured weight to zero despite the fact that t...

Page 96: ...a Select the size of the bag you attached b Tap OK Select Bag Type window sample North American version At the pump screen the Run button becomes a Pause button The compounder pumps each ingredient o...

Page 97: ...t before or during the compounding process refer to Fulfilling the Order Additional Steps on Page 91 If you need to stop compounding temporarily you can perform either of the following actions The Mix...

Page 98: ...ent Bag Always use aseptic technique when removing the bag 1 Remove the bag s fill port from the load cell s fill port holder 2 Remove the bag from the load cell 3 Clamp the bag s fill port 4 Disconne...

Page 99: ...auto addition allows you to add an ingredient to the existing configuration temporarily to fulfill the current order instead of selecting a new configuration which would require you to prime and veri...

Page 100: ...e it Refer to Removing the Patient Bag on Page 90 6 Attach a calibration bag Refer to Attaching the Calibration Bag on Page 73 7 Prime and verify the new inlet and ingredient Refer to Priming and Veri...

Page 101: ...is less than the 0 2 mL minimum required for use on the compounder If a formula loaded by scanning a barcode includes an ingredient that must be added manually A Manual Add button appears on the left...

Page 102: ...source container is appropriately depleted WARNING If there is a large amount of fluid remaining in the source container or if the container has emptied completely and forced air into the inlet there...

Page 103: ...a different container may require you to change and prime the inlet Refer to Attaching the New Ingredients and Inlets on Page 48 and Priming and Verifying on Page 58 Tap Stop filling and throw bag aw...

Page 104: ...mine the volume of fluid displaced b If more than one bubble is present evaluate each bubble and add the values together to determine the total volume of fluid displaced c Identify all the ingredients...

Page 105: ...ev C Page 97 7 At the Really abort the current solution message tap Yes Message The software unloads the formula 8 At the Operation Cancelled message tap OK Message 9 At the Fluid path will be flushed...

Page 106: ...ation on Page 132 Clean the channel over the bubble detector Refer to CLEANING THE COMPOUNDER on Page 101 Make sure that the outlet tube is in the proper position It should be at the bottom of the cha...

Page 107: ...Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 99 Message 3 At the pump screen tap Stop 4 At the Really abort the current solution message tap Yes Message The software unloads the formula 5 At t...

Page 108: ...g the Tube Set on Page 42 The occlusion detector is not damaged or dirty The outlet tube is straight and flat on the occlusion detector 7 At the Fluid path will be flushed with UI message tap OK Messa...

Page 109: ...y setup For instructions refer to Changing the Tube Set on Page 42 CLEANING THE COMPOUNDER Clean the compounder when indicated by your facility s protocol Use only these approved materials Non abrasiv...

Page 110: ...thumbscrew and washer used to attach the rotor Removing the thumbscrew 5 Remove the pump rotor from the spindle Removing the pump rotor 6 Using the approved materials clean the a Pump rotor making sur...

Page 111: ...re records important device activity in a Blackbox log which Baxter can access through the Blackbox Report when needed The compounder stores this data for a specified time period To set up the storage...

Page 112: ...se Backup All 2 At the Backup Database Location window a If desired change the location of the backup by tapping the button to the right of the current location not recommended b If you want to Replac...

Page 113: ...s Database Backup Blackbox 2 At the Backup Blackbox Location window a If desired change the location of the backup by tapping the button to the right of the current location not recommended b If you w...

Page 114: ...or the Baxter ExactaMix 2400 Compounder Page 106 53007410 Rev C 3 At the Backup succeeded message tap OK Message Compacting the Database 1 At the menu screen tap Tools Database Maintenance 2 At the Do...

Page 115: ...user groups and permissions refer to Setting Up the Users on Page 123 To change the Universal Ingredient for the most recently used configuration when you are not in the process of compounding a solut...

Page 116: ...ersal Ingredients refer to Using the Formulary Editor on Page 139 2 At the Change Universal window a Select the Universal Ingredient you want to use b Tap OK Change Universal window The compounder req...

Page 117: ...esent the actual volume remaining in the source container Change a remainder value only when you know the precise amount remaining in the source container Incorrect remainder values can lead to bubble...

Page 118: ...Performing Advanced Tasks Operator Manual for the Baxter ExactaMix 2400 Compounder Page 110 53007410 Rev C Edit Source Remainders window...

Page 119: ...to Setting Up the Users on Page 123 NOTE At any tab of the Options window Tapping OK saves the changes made on all the tabs and requires you to exit the software Tapping Cancel closes the window witho...

Page 120: ...uthorization if you want to require a qualified user to log in with a password to approve each MixCheck Report on the screen Select Enable Auto Display if you want the MixCheck Report to appear automa...

Page 121: ...hat the bag on the load cell may not be empty Tip Baxter recommends selecting Check for empty bag Track Product Expiration Date and Lot Number Select Enabled if you want to track the expiration date a...

Page 122: ...than the specified storage period is purged when the software starts up You can increase or decrease these settings For Solution Storage Days enter the number of days that used formulas are available...

Page 123: ...hnical Services Refer to Getting Help on Page 22 Select Enable if you want the compounder to display a message when the tube set has been used longer than recommended For Max Hours to Use Tube Set ent...

Page 124: ...inuation of the System tab To access the continued system options 1 At the menu screen tap Tools Options 2 At the Options window tap the System Cont tab Options window System Cont tab Database Compact...

Page 125: ...recommends entering 5 for the Final Solution and 5 for the Individual Ingredient Manual Add For Max manual add volume enter the maximum volume allowed for a manual addition If the volume of the formul...

Page 126: ...Page 118 53007410 Rev C Options window System Cont tab MixCheck Signature Label Enter text that you want to include at the bottom of the MixCheck Report Authorization Report Signature Label Enter text...

Page 127: ...General Select Use Security if you want each user to sign in with a user ID and password Select Remember last login if you want the login box to populate automatically with the user ID of the last us...

Page 128: ...ding formulas If you select this check box these options become available Select Use barcode verification if you want to require the use of a barcode reader for scanning labels during verification of...

Page 129: ...number of times a specific formula can be used for compounding For Max enter the maximum of times an individual formula can be used Tip Baxter recommends entering 1 for the Max SETTING UP THE DIRECTOR...

Page 130: ...on changes the locations of the directories from their current settings 4 If you want to change the locations of the directories a Tap Browse for Formula Files Reports or Backups b Select the location...

Page 131: ...ns to perform the required tasks WORKING WITH GROUPS Adding or Editing a Group IMPORTANT These functions require Administration permissions 1 At the menu screen tap Tools Users Edit Users and Groups E...

Page 132: ...v C Edit Group name window 3 At the Add Group window or Edit Group name window select the permissions for the group IMPORTANT These permissions will apply to an entire group of users You cannot assign...

Page 133: ...dient Change the Universal Ingredient without changing the configuration Administrator Pharmacist and Technician Verification Perform cosignature authorization of the priming and verifying steps durin...

Page 134: ...n requires Administration permissions 1 At the menu screen tap Tools Users Edit Users and Groups 2 At the Edit Users and Groups window in the top half a Select the group you want to delete from the Na...

Page 135: ...e menu screen tap Tools Users Edit Users and Groups The Edit Users and Groups window appears Edit Users and Groups window 2 At the Edit Users and Groups window a In the top half select the group from...

Page 136: ...gin Name b Enter the User Name Tip Baxter recommends using a short Login Name and full User Name c Select the Group to which the user is assigned d Tap Save 4 At the Edit Users and Groups window tap O...

Page 137: ...delete from the Login Name list then tap Delete 3 At the Delete user name message tap Yes to delete the user Message 4 At the Edit Users and Groups window tap OK Changing a Password 1 At the menu scr...

Page 138: ...ounder Page 130 53007410 Rev C Change Password window Logging in as a Different User 1 At the menu screen select Tools Users Change User 2 At the Login window a Enter a different Login name b Enter th...

Page 139: ...nd permissions refer to Setting Up the Users on Page 123 Before making any changes in the Configuration Editor contact Baxter Technical Services Refer to Getting Help on Page 22 IMPORTANT The Configur...

Page 140: ...placed into service Installers support services are available for reviewing and approving all new and updated configurations Refer to Getting Help on Page 22 for additional instruction 1 At the Edit...

Page 141: ...nfiguration Editor Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 133 Add Configuration window 2 Enter the Name of the configuration 3 Tap a port The Edit Port number win...

Page 142: ...is used during setup of the compounder This volume must be set for each product that is in each configuration Auto addition ingredients are always primed with the standard volume c If the selected pro...

Page 143: ...ow with products 6 At the configuration window a Select the Universal Ingredient NOTE For the Universal Ingredient to appear in the list it must first be in the configuration Universal Ingredients mus...

Page 144: ...formulary Refer to Using the Formulary Editor on Page 139 Selecting auto additions 7 To select auto additions a Select the check box for each desired ingredient b To select a specific product or add...

Page 145: ...wn in the sequence b When you are finished tap Return to Products 10 At the configuration window tap OK 11 At the Edit Configurations window tap Close WARNING A Baxter pharmacist must approve every ne...

Page 146: ...ons window a Select one or more configurations you want to delete b Tap Delete 2 At the Delete configuration name message tap Yes to delete the configuration Yes to All to delete all configurations se...

Page 147: ...gredients and products in the formulary IMPORTANT These functions require Formulary permissions For more information about user groups and permissions refer to Setting Up the Users on Page 123 Before...

Page 148: ...ars The top list identifies the ingredients that can be included in a formula The bottom list identifies each Product Name with its Drug ID that can be used for each ingredient type Tapping the Inlets...

Page 149: ...e Add Ingredient window or Edit Ingredient name window a Enter the Name b Enter the Abbr abbreviation NOTE The information entered in the Abbr field will appear on the Ingredient Abbreviation preview...

Page 150: ...Tap Delete 2 At the Delete ingredient name message tap Yes to delete the ingredient Yes to All to delete all ingredients selected in the Formulary Editor window NOTE If only one ingredient is selected...

Page 151: ...echnical Services Refer to Getting Help on Page 22 Viewing the Calibration Ingredient 1 At the Formulary Editor window in the top half tap Show Cal Ingredient The calibration ingredient is highlighted...

Page 152: ...NING It is important to select the correct inlet type for the container Selecting the incorrect inlet type can lead to occlusions and incorrect volume delivery resulting in patient harm b Select the a...

Page 153: ...e attached to the compounder The compounder displays a message if the product remains attached longer than the specified time g Select the product s Container Type h If you want to change the Name tap...

Page 154: ...ditor window a In the top half select the ingredient from the Name list b In the bottom half select the product you want to view from the Product Name list then tap Contained In A message with the pro...

Page 155: ...alcium members group and any phosphate containing products must be assigned to the phosphate members group to ensure the software will warn users about formulas that may cause a precipitate in the tub...

Page 156: ...edit or delete ingredient groups Calcium and Phosphate groups are created automatically Ingredients Groups window ADDING OR EDITING AN INGREDIENT GROUP 1 At the Ingredient Groups window do one of thes...

Page 157: ...t that is not listed as incompatible c Tap to move the group to the Incompatible Groups list To make another group compatible with this group a Select the other group in the Incompatible Groups list b...

Page 158: ...olumes for all the inlets IMPORTANT These functions require Administration permissions For more information about user groups and permissions refer to Setting Up the Users on Page 123 Before making an...

Page 159: ...OTE Use of this function is not recommended without Baxter instruction WARNING Modifying these settings without contacting Baxter Technical Services could result in ingredients over or under deliverin...

Page 160: ...Tap Delete 2 At the Delete inlet name message tap Yes to delete the inlet Yes to All to delete all inlets selected in the Inlet Editor window NOTE If only one inlet is selected only one will be delete...

Page 161: ...ical Services for adding updating flow factors IMPORTANT These settings affect the delivery volume so they must be accurate Do not make any changes unless directed by Baxter Technical Services Baxter...

Page 162: ...Compounder Page 154 53007410 Rev C 1 At the Inlet Editor window a Select an inlet b Tap Speeds Flow Factors 2 At the Speeds for Inlet name window follow the instructions of Baxter Technical Services...

Page 163: ...xter Technical Services Refer to Getting Help on Page 22 IMPORTANT These functions require Administration permissions For more information about user groups and permissions refer to Setting Up the Use...

Page 164: ...ntory Editor Operator Manual for the Baxter ExactaMix 2400 Compounder Page 156 53007410 Rev C The Bag Inventory window appears It lists the available bags and allows you to add edit or delete bags Bag...

Page 165: ...se options Tap Add to add a new bag The Add Bag window appears Select the bag you want to edit then tap Edit The Edit Bag name window appears Add Bag window 2 At the Add Bag window or Edit Bag name wi...

Page 166: ...BAG 1 At the Bag Inventory window a Select one or more bags you want to delete b Tap Delete 2 At the Delete bag name message tap Yes to delete the bag Yes to All to delete all bags selected in the Bag...

Page 167: ...8 cm paper or on A4 paper size of 70X25 4 mm or comparable paper NOTE For using A4 paper refer to Printing Options on Page 16 IMPORTANT Viewing printing and exporting reports requires Report permissio...

Page 168: ...and or the bottom The top of the report screen may include these navigation options o The print icon allows you to send the report to the specified printer o The export icon allows you to save the rep...

Page 169: ...Report To make this report display and or print automatically after compounding and or allow on screen authorization refer to MixCheck Report on Page 112 To require a cosigner to authorize the MixChec...

Page 170: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder Page 162 53007410 Rev C Sample MixCheck Report...

Page 171: ...on the patient bag and or the original formula 2 Add these ingredients to the solution manually 3 Place your initials in the Added column Evaluate the information in the Details section to verify tha...

Page 172: ...he Measured Weight of the solution to the Expected Weight Verify that the Difference is within acceptable limits per your facility s protocol Verify that each ingredient and its Ordered Volume match t...

Page 173: ...v C Page 165 If the Serial Number of the formula file contains more than 36 characters then the MixCheck Report prints up to 36 characters followed by Sample MixCheck Report If the Serial Number of th...

Page 174: ...rts 1 At the menu screen tap Reports MixCheck Old MixCheck Data 2 At the Solutions window a Select a formula b Tap one of these buttons View MixCheck to view the report on the screen Print MixCheck to...

Page 175: ...be set up to populate these columns automatically The ingredient name port and inlet used during setup The following options are available for customizing the Authorization Report To require a cosigne...

Page 176: ...ific formula To view the Formula Report 1 At the menu screen tap Reports Standard Formula 2 At the Select Formula window a Select Show All Formulas or select another filter to reduce the number of for...

Page 177: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 169 Sample Formula Report...

Page 178: ...3007410 Rev C If the Serial Number of the formula file contains more than 50 characters then the Formula Report prints up to 50 characters followed by Sample Formula Report If the Serial Number of the...

Page 179: ...Page 171 LOG REPORT The Log Report is a summary of the formulas that were used for compounding on a specific day To view the Log Report 1 At the menu screen tap Reports Standard Log 2 At the Select Re...

Page 180: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder Page 172 53007410 Rev C Sample Log Report...

Page 181: ...10 Rev C Page 173 If the Serial Number of the formula file contains more than 80 characters then the Log Report prints up to 80 characters followed by Sample Log Report If the Serial Number of the for...

Page 182: ...v C CONFIGURATION REPORT The Configuration Report provides information about a specific configuration To view the Configuration Report 1 At the menu screen tap Reports Standard Configuration 2 At the...

Page 183: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 175 Sample Configuration Report...

Page 184: ...53007410 Rev C FORMULARY REPORT The Formulary Report lists the ingredients that are included in the formulary Only ingredients that may actually be mounted on the compounder should be in the formulary...

Page 185: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 177 Sample Formulary Report...

Page 186: ...ort displays the barcodes for products that are in the formulary You can print the report onto labels for use with products that do not have a manufacturer s barcode To view the Product Barcodes Repor...

Page 187: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 179 Sample Product Barcodes Report...

Page 188: ...RCODES REPORT The Inlet Barcodes Report displays the barcodes for the inlets You can print the report onto labels in case you make a mistake during setup and need an extra barcode label To view the In...

Page 189: ...ist of all important system activity for a specific period of time If necessary Baxter may use this information for troubleshooting To view the Blackbox Report 1 At the menu screen select Reports Stan...

Page 190: ...s Operator Manual for the Baxter ExactaMix 2400 Compounder Page 182 53007410 Rev C For assistance with reading this report contact Baxter Technical Services Refer to Getting Help on Page 22 Sample Bla...

Page 191: ...eriod of time It also distinguishes between automatic and manual calibrations To view the Calibration Summary Report 1 At the menu screen tap Reports Standard Calibration Summary 2 At the Select Repor...

Page 192: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder Page 184 53007410 Rev C Sample Calibration Summary Report...

Page 193: ...f time To view the Formula Precision Report 1 At the menu screen tap Reports Standard Formula Precision 2 At the Select Report Dates window a Enter the First date and Last dates for the report NOTE Th...

Page 194: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder Page 186 53007410 Rev C Sample Formula Precision Report...

Page 195: ...187 If the Serial Number of the formula file contains more than 36 characters then the Formula Precision Report prints up to 36 characters followed by Sample Formula Precision Report If the Serial Num...

Page 196: ...Usage Report summarizes the ingredient usage during a specific period of time This report is used to manage inventory To view the Ingredient Usage Report 1 At the menu screen select Reports Standard I...

Page 197: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 189 Sample Ingredient Usage Report...

Page 198: ...Bag Usage Report summarizes the bag usage during a specific period of time This report is used to manage inventory To view the Bag Usage Report 1 At the menu screen tap Reports Standard Bag Usage 2 A...

Page 199: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 191 Sample Bag Usage Report...

Page 200: ...ted by Baxter Technical Services Sample Flow Factors Report Flow Factors Configuration TPN TEST Date 1 8 2020 Time 4 18 50PM Port Seq Product Name Min Vol mL Max Vol mL Flow Factor 6 1 Baxter K Chlori...

Page 201: ...e reader does not indicate that it is operational 1 Check that the green LED on the barcode reader illuminates when you scan a barcode If the LED does not illuminate check that the cable for the barco...

Page 202: ...s not meet the Order entry specification 1 Correct the serial number of the 2D Formula Barcode then print and scan the new 2D Formula Barcode Text The following formula name or data is not supported C...

Page 203: ...Barcode X cannot be loaded 1 Following manual Ingredient Name s exceeds the maximum characters 25 s You must correct the manual Ingredient Name s then load the formula The 2D Formula Barcode contains...

Page 204: ...arcode Text Barcode reader X barcode contains an invalid Code 39 character The barcodes for formulas must use the Code39 symbology which has a restricted character set 1 Reprint the barcode 2 Scan the...

Page 205: ...add volume of max manual add volume mL The ordered volume of the product exceeds the maximum volume for manual additions 1 Tap Cancel 2 Check that the Drug ID is correct and that it matches the code n...

Page 206: ...rn it right side up to prevent fluid from flowing 2 Remove the inlet from the incorrect port and attach it to the correct port 3 Return the source container to the vial rack or hanger Text Flushing wi...

Page 207: ...ormation about user groups and permissions refer to Setting Up the Users on Page 123 Tap one of these options Tap Change UI To to change the Universal Ingredient to one that has the minimum volume req...

Page 208: ...product and associated Inlets Contact Baxter Technical Services for help All ingredients need flow factors prior to compounding 1 Click Ok on the error message 2 Contact Baxter Technical Services for...

Page 209: ...his operation must be completed prior to pumping a solution Would you like to calibrate the pump now The pump is not calibrated 1 Tap Yes 2 Calibrate the pump Refer to Calibrating the Compounder on Pa...

Page 210: ...s out of range After the compounder delivers an ingredient the weight of the patient bag differs from the expected weight by more than the acceptable difference The compounder checks the weight after...

Page 211: ...cts that the destination bag is not attached If the bag is not attached attach the appropriate bag to the load cell Refer to Attaching the Patient Bag on Page 85 Attaching the Calibration Bag on Page...

Page 212: ...t and then exit the PRIME AND VERIFY screen The Occlusion Detector Test failed Pump door open Select OK to Retry Select Cancel to Exit The test did not finish because the pump door was opened during t...

Page 213: ...label for the patient bag cannot be printed Contact Baxter Technical Services Refer to Getting Help on Page 22 The MixCheck Report does not print The printer is disconnected or turned off Check that t...

Page 214: ...A software error occurred Reconnect the cord and cables Refer to Step 6 on Page 24 Text No Valve Device Assigned A software error occurred Reconnect the cord and cables Refer to Step 6 on Page 24 Tex...

Page 215: ...g Help on Page 22 USB drive NOTE Be sure that the USB drive is free of viruses Barcode reader at the compounder In the order entry software the pharmacist creates an order which creates a barcode and...

Page 216: ...xactly match the code number for that product in the compounder s formulary If a code number is assigned to one product in the order entry software and that number is assigned to a different product i...

Page 217: ...to backup message if you Want to back up the current database before restoring tap Yes and continue with the next step Do not want to back up the current database before restoring tap No and skip to s...

Page 218: ...t the Backup succeeded message tap OK Message 5 At the Restore Database Location window a If desired change the location of the backup you want to restore by tapping the button to the right of it b Ta...

Page 219: ...the compounder s components sit Blackbox data The logged activities of the compounder mainly based on the communication between software firmware and user actions cardioplegia A specific combination...

Page 220: ...ainer types or manufacturers final flush A delivery of fluid that is pumped to clear all delivered ingredients from the common fluid pathway to ensure that these ingredients are fully present in the f...

Page 221: ...creen window or other text that provides information or instructions but does not allow you to enter information Pay attention to the text because some messages may include critical information micro...

Page 222: ...port a patient s nutritional needs tube set The valve set and inlets Universal Ingredient UI The ingredient that is used to flush the common fluid pathway This ingredient must be contained in the conf...

Page 223: ...supporting USB 1 1 2 0 The display supports the connection and use of a USB 1 1 keyboard and mouse Dimensions Main module Length 24 in 61 cm Width 10 in 25 4 cm Height 10 in 25 4 cm Display Length 11...

Page 224: ...47 F 0 64 C Maximum relative humidity 10 80 Maximum altitude Not to exceed 3 000 m Main supply voltage fluctuation Not to exceed 10 Sound pressure level Not to exceed 85 dBA For Indoor Use Only ISO Cl...

Page 225: ...equipment is under warranty Baxter will replace the defective equipment Equipment that is not under warranty can also be replaced however the customer is responsible for the cost of repairs and shipp...

Reviews: