Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C
Page 13
FEATURES
SUMMARY OF FEATURES
The compounder:
•
Accepts formulas created by order-entry software via 2D Formula Barcode/PAT/FRM file
interface, or by direct entry on the compounder
•
Uses barcodes on the source containers and inlets to promote proper setup
•
Includes software with a Setup Wizard to guide you through the setup process
•
Supports a maximum of 24 ingredients, source containers in volumes of 10–5,500 mL
and destination bags in volumes of 125–5,000 mL
•
Allows you to attach the same ingredient to more than one port, creating an electronic
Y-site
•
Allows you to specify the sequence in which ingredients are pumped
•
Allows you to specify the accuracy limits for the finished solution
•
Uses volumetric delivery, gravimetric verification and automatic calibration to help
ensure delivery accuracy
•
Uses a bubble detector and occlusion detector
•
Can be immediately stopped by lifting the pump door
•
Can track ingredient lot numbers and ingredient expiration dates
•
Generates a MixCheck™ Report for each finished solution
•
Can print reports and barcode labels at the compounder’s printer
•
Can be set up to communicate with the printer and the order-entry computer through a
network
ORDER ENTRY
Order entry can be done through direct entry, refer to page 77, or by using separate order-
entry software.
The ExactaMix operating software can communicate, via a network, with order-entry software
on a separate computer. The order-entry software must produce both a .PAT/.FRM file and a
corresponding barcode. Scanning the barcode at the compounder retrieves the .PAT/.FRM file.
Alternatively the order-entry software must be able to produce a formula label, containing the
formula details in the 2D barcode. Scanning the 2D barcode loads the formula onto the
compounder. For more information, refer to Loading the Formula on Page 75.
FORMULARY
The formulary is the list of ingredients, and associated products, which can be attached to the
compounder.
Summary of Contents for ExactaMix 2400 Compounder