Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C
Page 109
5.
When the
Completed flushing
message appears, remove the calibration bag. Refer to
Removing the Calibration Bag on Page 73.
6.
If you will continue with compounding a solution, attach a patient bag. Refer to
Attaching the Patient Bag on Page 85.
7.
At the
Completed flushing
message, tap
OK
.
Message
CHANGING THE INGREDIENT REMAINDERS
For each ingredient, the compounder tracks the volume that is used and the volume that
remains in the source container (the remainder). If necessary, you can manually change each
remainder shown in the software.
WARNING
The remainder value in the software must accurately represent the actual volume
remaining in the source container. Change a remainder value only when you know the
precise amount remaining in the source container. Incorrect remainder values can
lead to bubbles, occlusions and under-delivery of an ingredient if its source container
runs empty.
IMPORTANT!
This function requires Compounder permissions. For more information about
user groups and permissions, refer to Setting Up the Users on Page 123.
1.
At the menu screen, tap
Compound > Edit Source Remainders
.
2.
At the
Edit Source Remainders
window:
a.
For the desired ingredient, change the
Remainder
to correspond to the volume
remaining in the container.
NOTE:
You can tap
Reset
to change the remainder to its default value, or tap
Reset All
to change all the remainders to their default values at the same time.
Tip!
Baxter does not recommend using
Reset All
unless the entire list of
ingredients has been changed.
b.
Tap
OK
.