Issues with the Formulas, Ingredients and Configurations
Issue / On-screen Text
Explanation
Suggested Actions
Text:
This formula contains
<
ingredient name
>, which
is not currently on the
configuration. Do you
wish to add it to the
configuration?
The ingredient is an auto-
addition ingredient.
Add the ingredient. Refer to Performing an Auto-
Text:
Configuration must be
verified before
compounding.
The configuration is not
primed and verified.
Prime and verify the configuration. Refer to Priming
and Verifying on Page 58.
Text:
The following ports
contain products that
have been spiked longer
than allowed. Port
Product Time Spiked
Allowed Hang Time
(Hours) <
list of products
>.
One or more ingredients
are expired.
1.
Tap
Cancel
.
2.
Check the expiration dates of all the
ingredients.
3.
Change the container of each expired
ingredient.
Text:
Ingredient <
ingredient
name
> must be manually
added and its requested
volume of <
requested
volume
> mL exceeds the
maximum manual add
volume of <
max manual
add volume
> mL.
The ordered volume of the
product exceeds the
maximum volume for
manual additions.
1.
Tap
Cancel
.
2.
Check that the
Drug ID
is correct and that it
matches the code number from the order-entry
software. Refer to Adding or Editing a Product
on Page 143.
Text:
Manual Add
The product is not
included in the
configuration, or its
ordered volume is less
than 0.2 mL.
Add the product manually. Refer to Performing a
Manual Addition on Page 93.
The product is included in
the physical configuration,
and its ordered volume is
at least 0.2 mL, but it does
not match any products in
the software
configuration.
Check that the
Drug ID
is correct and that it matches
the code number from the order-entry software.
Refer to Adding or Editing a Product on Page 143.
Text:
Swap Container: Your
container of <
ingredient
name
> is empty. You
have <
remaining volume
>
mL left to run. Please
change the container
now.
The container is almost
empty and needs to be
replaced.
Check that the container is almost empty.
•
If the container is almost empty, replace it.
Refer to Replacing a Source Container on
Page 94.
•
If the container is not almost empty, check that
it is the correct container, and contact Baxter
Technical Services. Refer to Getting Help on
Page 22.