Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C
Page 37
CALIBRATING THE LOAD CELL
This procedure calibrates the load cell to ensure that it measures weight accurately.
The load cell must be calibrated:
•
Daily when the calibration expires
•
Using a 2,000 g weight that Baxter provides
Tip!
Baxter recommends:
•
Using gloves to handle the weight, to minimize the accumulation of oils and dust
•
Storing the weight in its storage case
To calibrate the load cell:
1.
At the
Setup Wizard
screen, tap
Calibrate Load Cell
.
Setup Wizard screen, calibrating the load cell
NOTE:
To calibrate the load cell without using the Setup Wizard, you can tap
Compound
> Calibrate Load Cell
at the menu screen.