USER MANUAL BATMODE 2S/2S+ SYSTEM
54
directory. For both, single-channel and multi-channel data acquisition, the filenames of the recorded
data are always composed as follows:
[ID of BATmode][date and time of acquisition][autoincremented number].WAV
If you use multiple UltraSoundGates 116Hnbm/116Hnbm-r, please note that the
channel-to-UltraSoundGate assignment will not change unless the UltraSoundGates
are disconnected from the BATmode system. If you have disconnected the
UltraSoundGates from the system, make sure that after reconnection a certain
channel also represents the corresponding measurement point.
By using the button <L
oad default RECORDER settings
> you can load default BMU-configuration for
RECORDER software. These default BMU-configuration conforms to the settings recommended by bat
bioacoustictechnology. On loading default values current settings will be discarded.
Please consider that all settings, which are employed within the RECORDER software,
are saved on closing the software.
Also running the RECORDER software externally, that means, it is not started by
BATcontrol, may affect settings, if one of the configuration files (depends on the
selected number of measurement channels)
-
“
C:\Users\BAT\AppData\Roaming\BATcontrol\
recorderconfig.ini“,
-
“
C:\Users\BAT\AppData\Roaming\BATcontrol\
recorderconfig1.ini“,
-
“
C:\Users\BAT\AppData\Roaming\BATcontrol\
recorderconfig2.ini“,
-
“
C:\Users\BAT\AppData\Roaming\BATcontrol\
recorderconfig3.ini“
or
-
“
C:\Users\BAT\AppData\Roaming\BATcontrol\
recorderconfig4.ini“
is loaded.
To change settings of the RECORDER-Software, which are not covered by BATcontrol,
we recommend the following approach:
1)
Start monitoring in BATcontrol
2)
When the RECORDER-Software is opened, stop the recording by
pressing the Stop-Button in the RECORDER-Software
3)
Change settings in the RECORDER-Software
4)
Close the RECORDER-Software
5)
Restart monitoring with BATcontrol
Please note that the settings made outside of BATcontrol are only saved for the current
number of measurement channels.