Troubleshooting
10. Alarm sounds and screen displays:
ALARM: TANK NAME
TOO LONG TO FILL
Printer records:
* ALARM
TANK NAME
TOO LONG TO FILL
HH:MM:SS
YY/MM/DD
(Tank Name) water level not reached within specified
time.
1.
Press
Silence Buzzer
or
ACK
touch pad to silence
alarm buzzer.
2.
Building water supply valves not fully open or closed –
Verify and open building supply valves as necessary.
3.
Building air supply too low – Call STERIS.*
4.
Tank filling alarm time too short (4 min. default) – Call
STERIS.*
5.
Not enough building water pressure – Verify or adjust
building water pressure.
6.
(Tank Name) outlet valve open – Call STERIS.*.
11. Alarm sounds and screen displays:
ALARM: TANK NAME
TOO LONG TO DRAIN
Printer records:
*ALARM:
TANK NAME
TOO LONG TO DRAIN
HH:MM:SS YY/MM/SS
(Tank Name) water level sensor is still activated within
specified time or less than 100 L detected when filling
tank.
1.
Press
Silence Buzzer
or
ACK
touch pad to silence
alarm buzzer.
2.
Building water supply valves closed – Open air supply
valves.
3.
Building water supply pressure too low – Call STERIS.*
4.
(Tank Name) outlet valve always closed – Call STERIS.*
5.
(Tank Name) filling valve always open – Call STERIS.*
* Service charges may be incurred. Consult your warranty for details.
Table 7-4. Troubleshooting Chart - Alarm and/or Printout (Cont’d)
ALARM MESSAGE
POSSIBLE CAUSE AND CORRECTION