Maintenance
79
03/2016
CFPP Process Analyzer CFPP-4.2
10.10 Changing the battery of the box PC
The analyzer is controlled by a box PC. It is located in the control box (see page Fehler!
Textmarke nicht definiert.). The box PC is equipped with a Lithium battery. During voltage
breakdown it ensures among other things, that the system clock carries on running. To
ensure the functionality of the box PC, you should change the battery every five years.
WARNING!
Danger of explosion by changing the battery in potentially explosive
atmosphere!
Dismounting and installing the battery can cause sparks. They can ignite
potentially explosive atmosphere!
FOR THIS REASON:
-
Make sure, that during all operations for changing the battery no
potentially explosive atmosphere is present.
For changing the battery you must open the control box.
•
For it switch off the analyzer (see page 65).
•
Make sure, that the power supply is switched off.
•
Open the control box.
WARNING!
Danger of explosion by using unsuitable batteries in potentially explosive
atmosphere!
Unsuitable batteries can ignite potentially explosive atmosphere by
overheating in the event of an error!
FOR THIS REASON:
-
For replacement use only batteries from BARTEC BENKE.
Summary of Contents for CFPP-4.2
Page 1: ...CFPP Process Analyzer CFPP 4 2 Operating Instructions 395778MDHEN A 03 2016 ...
Page 6: ...IV Contents CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 7: ...General 5 03 2016 CFPP Process Analyzer CFPP 4 2 1 General ...
Page 9: ...Safety 7 03 2016 CFPP Process Analyzer CFPP 4 2 2 Safety ...
Page 25: ...Introduction 23 03 2016 CFPP Process Analyzer CFPP 4 2 3 Introduction ...
Page 28: ...26 Introduction CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 29: ...Technical Data 27 03 2016 CFPP Process Analyzer CFPP 4 2 4 Technical Data ...
Page 33: ...Analyzer Description 31 03 2016 CFPP Process Analyzer CFPP 4 2 5 Analyzer Description ...
Page 46: ...44 Transport Packaging and Storage CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 55: ...Initial Operation 53 03 2016 CFPP Process Analyzer CFPP 4 2 8 Initial Operation ...
Page 64: ...62 Initial Operation CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 65: ...Operation 63 03 2016 CFPP Process Analyzer CFPP 4 2 9 Operation ...
Page 69: ...Maintenance 67 03 2016 CFPP Process Analyzer CFPP 4 2 10 Maintenance ...
Page 102: ...100 Maintenance CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 103: ...Troubleshooting 101 03 2016 CFPP Process Analyzer CFPP 4 2 11 Troubleshooting ...
Page 107: ...Troubleshooting 105 03 2016 CFPP Process Analyzer CFPP 4 2 ...