102
Troubleshooting
CFPP Process Analyzer CFPP-4.2 03/2016
In case of a malfunction, first check if the PACS process software shows any error
messages. In the software manual, you will find a list of error messages, possible causes
and suggested solutions.
If the process software does not display any error messages, the following list helps you to
identify the problem and its possible causes.
If you are unable to solve the problem, we will be pleased to help you:
BARTEC BENKE GmbH
Borsigstraße 10
21465 Reinbek/Hamburg
Germany
Tel.:
+49 (0) 40 -72703 0
Fax:
+49 (0) 40 -72703 363
E-Mail: [email protected]
Error
Possible cause
Troubleshooting
Analyzer is switched off /
does not start
a) Display of overpressure
control is not present
Power supply of the analyzer
is interrupted (e.g. broken or
blown fuse)
Check and restore power
supply
Fusing of overpressure
control system is faulty.
Remove the cause of the
malfunction.
Analyzer is switched off /
does not start
a) Display of overpressure
control is present
Overpressure control system
has switched off the analyzer
/ cannot start the analyzer
Please refer to operating
manual of the overpressure
control system.
Internal fuses are faulty /
blown
Remove the cause of the
malfunction.
Power supply of the
I/O board interrupted or
I/O board faulty
Remove the cause of the
malfunction.
Instrument air supply is
interrupted
Remove the cause of the
malfunction.
No Ready signal /
measuring result over a
longer period of time
Analyzer is switched off.
see above
PACS is in “Standby” or
“Shutdown” state; or PACS is
not operating;
Please refer to the software
manual.
Unsuitable parameter settings Adjust the parameters (see
software manual).
Cable connection to the
analyzer is interrupted
Remove the cause of the
malfunction.
Measuring value output
signals an unusual or
Sudden significant changes in
product properties
Summary of Contents for CFPP-4.2
Page 1: ...CFPP Process Analyzer CFPP 4 2 Operating Instructions 395778MDHEN A 03 2016 ...
Page 6: ...IV Contents CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 7: ...General 5 03 2016 CFPP Process Analyzer CFPP 4 2 1 General ...
Page 9: ...Safety 7 03 2016 CFPP Process Analyzer CFPP 4 2 2 Safety ...
Page 25: ...Introduction 23 03 2016 CFPP Process Analyzer CFPP 4 2 3 Introduction ...
Page 28: ...26 Introduction CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 29: ...Technical Data 27 03 2016 CFPP Process Analyzer CFPP 4 2 4 Technical Data ...
Page 33: ...Analyzer Description 31 03 2016 CFPP Process Analyzer CFPP 4 2 5 Analyzer Description ...
Page 46: ...44 Transport Packaging and Storage CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 55: ...Initial Operation 53 03 2016 CFPP Process Analyzer CFPP 4 2 8 Initial Operation ...
Page 64: ...62 Initial Operation CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 65: ...Operation 63 03 2016 CFPP Process Analyzer CFPP 4 2 9 Operation ...
Page 69: ...Maintenance 67 03 2016 CFPP Process Analyzer CFPP 4 2 10 Maintenance ...
Page 102: ...100 Maintenance CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 103: ...Troubleshooting 101 03 2016 CFPP Process Analyzer CFPP 4 2 11 Troubleshooting ...
Page 107: ...Troubleshooting 105 03 2016 CFPP Process Analyzer CFPP 4 2 ...