![background image](http://html1.mh-extra.com/html/bartec-benke/cfpp-4-2/cfpp-4-2_operating-instructions-manual_3949374021.webp)
Safety
19
03/2016
CFPP Process Analyzer CFPP-4.2
The analyzer is constructed for operation in a explosive atmosphere and contains potential
ignition sources.
WARNING!
Possible explosion or fire!
An explosion or a fire can cause serious injury or death. FOR THIS REASON:
-
Do not open the housing while a potentially explosive atmosphere
may be present, or while the analyzer is under electrical power.
-
After switching off the power to the analyzer, wait for at least 5
minutes before you open the control box (to allow hot surfaces to
cool down and electric charges to decrease).
-
If you want to open an Ex housing with the power switched on:
Make sure that there is no explosive atmosphere.
-
If you want to open an Ex housing with the power switched on: First
get a hot-work permit from the operating company.
-
After finishing work, close and lock the Ex housing (explosive
atmosphere may develop later).
-
Never clean plastic parts (such as the touchscreen) with a dry cloth,
but always use a damp cloth to avoid static electricity discharge and
sparks.
-
Do not remove or disable any protective devices.
Summary of Contents for CFPP-4.2
Page 1: ...CFPP Process Analyzer CFPP 4 2 Operating Instructions 395778MDHEN A 03 2016 ...
Page 6: ...IV Contents CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 7: ...General 5 03 2016 CFPP Process Analyzer CFPP 4 2 1 General ...
Page 9: ...Safety 7 03 2016 CFPP Process Analyzer CFPP 4 2 2 Safety ...
Page 25: ...Introduction 23 03 2016 CFPP Process Analyzer CFPP 4 2 3 Introduction ...
Page 28: ...26 Introduction CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 29: ...Technical Data 27 03 2016 CFPP Process Analyzer CFPP 4 2 4 Technical Data ...
Page 33: ...Analyzer Description 31 03 2016 CFPP Process Analyzer CFPP 4 2 5 Analyzer Description ...
Page 46: ...44 Transport Packaging and Storage CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 55: ...Initial Operation 53 03 2016 CFPP Process Analyzer CFPP 4 2 8 Initial Operation ...
Page 64: ...62 Initial Operation CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 65: ...Operation 63 03 2016 CFPP Process Analyzer CFPP 4 2 9 Operation ...
Page 69: ...Maintenance 67 03 2016 CFPP Process Analyzer CFPP 4 2 10 Maintenance ...
Page 102: ...100 Maintenance CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 103: ...Troubleshooting 101 03 2016 CFPP Process Analyzer CFPP 4 2 11 Troubleshooting ...
Page 107: ...Troubleshooting 105 03 2016 CFPP Process Analyzer CFPP 4 2 ...