7.
Disconnect the membrane switch lead from the
printed circuit board. (See Figure 6.)
8.
Disconnect the resistivity cell lead from the printed
circuit board.
9.
Disconnect the printed circuit board relay plug by
squeezing the retaining clip while pulling the plug
straight out from the frame.
10. Remove the nuts holding the printed circuit board in
place.
11. Remove the printed circuit board. Disconnect power
supply (3 wires) from printed circuit board.
12. Reconnect power supply to new printed circuit
board and install the new printed circuit board.
Refer to electrical diagrams.
13. Reconnect the resistivity cell lead and printed circuit
board relay plug.
14. Reconnect the membrane switch lead.
15. Reinstall the front cover.
16. Retape the Luer fitting with new Teflon
®
tape and
reinstall fitting.
17. Reinstall the EASYpure RF’s cover.
18. Reconnect the power cord to the unit and the unit
to the power supply.
M
AINTENANCE AND
S
ERVICING