16
10. Install and rinse new cartridges according to the in-
structions in the Cartridge Installation section.
Do
not reinstall used cartridges or 0.2 micron filter
(they
may contain large amounts of bacteria.)
11. Reconnect to the power supply and press the
“START” button.
Fuse Replacement
1.
Disconnect the EASYpure RF from the power supply
and remove the reservoir and cap assembly.
2.
Remove the power cord from the power entry module.
3.
Remove the screws securing the EASYpure RF’s cov-
er.
4.
Remove the cover by lifting it straight up.
5.
Pull out the fuse drawer located in the power entry
module on the back left-hand side of the unit.
6.
Remove old fuses and replace with fuses of the same
type and rating. (See Parts Listing, page 18.)
7.
Replace fuse drawer.
8.
Replace cover and power cord and reservoir cap as-
sembly.
9.
Reconnect unit to power supply.
Printed Circuit Board Replacement (See Fig-
ure 6)
1.
Disconnect the unit from the power supply.
2.
Remove the power cord from the power entry module.
3.
Remove the reservoir cap and the screws securing
the EASYpure RF’s cover.
4.
Remove the cover by lifting it straight up.
5.
The filter and the Luer fitting.
6.
Remove the screws securing the front cover. Remove
the front cover.
M
AINTENANCE AND
S
ERVICING