11
Normal Operation
1.
Fill the feedwater reservoir with water and replace
the cap.
2.
Turn main power on at power entry module.
3.
Press the “START” button on the front of the
EASYpure RF.
The EASYpure RF’s pump will begin to run and the
Purity meter will display the resistivity of the water in
megohm-cm.
4.
Allow the water’s resistivity to rise to the desired
purity before drawing off water.
Water Draw off
1.
Remove the protective cap from the filter bell.
2.
Depress the draw-off lever.
3.
When draw off is complete, lift the draw-off lever
and replace the protective cap on the filter bell.
Feedwater Replenishment
As water is drawn-off from the EASYpure RF, the feed-
water reservoir will require refilling. To refill the reservoir:
1.
Remove the Ventgard cap. Do not allow water to
enter the Ventgard cap.
2.
Pour pretreated feedwater into the feedwater reser-
voir until the water is level with the top of the sight
gauge.
3.
Replace the Ventgard cap.
4.
Allow the water’s resistivity to rise to the desired
purity before drawing off water.
NOTE
For more demanding applications where low TOC
water is required, a third reservoir volume rinse of
the cartridges and filter may be necessary.
NOTE
On initial startup, the purity meter may display
“ERR.” This is caused by air in the cell and should
be replaced by a resistivity reading almost imme-
diately. If “ERR” does not go out after the pump
has run for a minute or if it appears any time while
the EASYpure RF is in operation, refer to the
Troubleshooting section of this manual.
Normal Operation