6
Specifications
Dimensions and Clearance requirements.
EASYpure RF dimensions - 12" W X 19" D X 18 1/8" H (30.5 cm X 48.3 cm X 46.0 cm).
Clearances:
Sides - 4" (10.1 cm) minimum to allow air flow
Above - 12" (30.5 cm) minimum for reservoir replenishment
Cartridge replacement requires that you be able to access the back of the unit and open the cartridge
access door (total depth, unit + open door, = 34") (86.4 cm).
Storage Reservoir Capacity - Approximately 6.5 liters usable, 7.0 liters total
Feedwater Requirements
The EASYpure RF requires water pretreated by either distillation, deionization or reverse osmosis.
TOC - Less than 1.0 ppm.
Turbidity - Less than 1.0 N.T.U.
Temperature - 40°F - 120°F (4.4°C - 48.9°C)
Resistivity (Minimum):
Distilled - 300,000 Ohm-cm
Deionized - 1.0 Megohms-cm
Reverse osmosis - 100,000 Ohm-cm
Product Water
Quality
Resistivity: ASTM Type I
TOC: Less than 5.0 PPB
Flow Rate: 0.5 LPM with a new final filter
Electrical Requirements
The EASYpure RF is equipped with a power cord to be plugged into an electrical outlet of the appropriate volt-
age.
Model D7032 - 100 VAC +5% -10%, 47-63 Hz.
Model D7031 - 120 VAC +5% -10%, 47-63 Hz.
Model D7033 - 240 VAC +5% -10%, 47-63 Hz.
Model D7033-33 - 230 VAC +10% -10%, 47-63 Hz.
Environmental Conditions
Operating: 4°C - 49°C; 20% to 80% relative humidity, non-condensing. Installation Category II (over-
voltage) in accordance with IEC 664. Pollution Degree 2 in accordance with IEC 664.
Altitude limit: 5,000 meters.
Storage: -25°C to 65°C; 10% to 85% relative humidity.
Specifications