12
Run and Standby Modes
Since not all qualities of permissible feedwater will reach
maximum resistivity after one pass through the unit’s car-
tridges (especially as the cartridges near exhaustion), the
EASYpure RF has two operational modes.
In the run mode, the pump continuously recirculates water
through the cartridges. This is the mode that the unit enters
upon startup. If water will be drawn from the EASYpure RF
on a continuous basis throughout the day, it is recommend-
ed that the EASYpure RF be left in the run mode. This will
ensure that the feedwater added to refill the EASYpure
RF’s reservoir will be up to maximum purity when it is
needed. In the run mode, the purity meter display indicates
the resistivity of the water available for draw off.
In standby mode, the pump runs for 10 minutes out of
every hour (i.e. ten minutes on, fifty minutes off). If water
will be drawn from the unit only infrequently, it is recom-
mended that the unit be put into standby mode. To put the
unit into standby mode, press the “STANDBY” button on
the front of the unit. The purity meter’s digital display will
display “Sby” to indicate that the unit is in standby mode.
N
ORMAL
O
PERATION