56-101-01175 Revision P Page | 20
Note
(*1)
:
When a remote iOS or Android device is connected to DPS1000, the touch screen on
the DPS1000 is disabled except for
the “Disconnect Remote” and “Local UI: enable”
APPENDIX E: WI-FI CONNECTION PROCEDURE
Note
(*2)
:
Wi-Fi is a free, embedded, secure, factory-enabled capability.
Note
(*3)
:
ADTS App is NOT supported when UUT is in Calibration Mode
Note
(*4)
:
The DPS1000 must be connected to an outlet with ground connection.
CAUTION:
THE DPS1000 SHOULD NEVER BE POWERD OFF AND RAPIDLY POWERED
BACK ON WITH NO WAITING TIME AS THIS MAY DAMAGE UNIT.
2. OPTIONS
It is important, whether testing on the aircraft or on the bench, that the Test Set be positioned
to not impede the ability to quickly power down and/or disconnect electrical power should the
need arise.
3. ACCESSORIES
A.
Included Items
1. Instruction Manual, P/N 56-101-01175 (refer to Figure 13).
2. Electrical Cable, P/N AE9887-ND (refer to Figure 14).
Key to Figure 12 (Continued)
5
Ground (Earth) Stud
Used to ensure Test Set chassis is at same
ground potential as the Aircraft or Instrument
being tested.
6
Fused Power Entry
Module
Connects external power to the tester. Unit uses
either a cable with a standard 115 VAC plug or the
optional cable for operator installation of 230 VAC
plug.
(*4)
7
Power Entry Fuse Holder
Power entry module fuses.
•
For 120 VAC operation use 1(ea) fuse: 4A.
•
For 220 VAC operation use 1(ea) fuse: 2A.
8
Power ON/OFF Switch
Used to power ON/OFF Test Set operation.
9
USB Port
Standard USB port used for multiple purposes:
•
To Upgrade Test Set HUIM or PCM
Firmware from flash drive (refer to
APPENDIX H: FIELD FIRMWARE
UPDATE PROCEDURE VIA USB &
BARFIELD ADTS APP
•
To download Test Set history log contents
onto USB flash drive.
10
Micro USB Port
OEM use only.