NioWatch operation
System manual E1
75
•
To adjust display luminance manually, adjust the
Luminance
slider. This function is not available for MFCD/MFGD 1218
displays.
The luminance setting is saved when the system is shut
down.
If the slider is positioned above the CAL mark, the luminance
is in calibrated position, as determined during the latest
calibration. If the slider is not above the CAL mark, the
luminance is not in calibrated position.
Important
If you want the system to be DICOM compliant, you must select
the DICOM display function and calibrate the display. After
calibration, the luminance must remain in calibrated position to
maintain DICOM compliance.
Summary of Contents for System manual E1
Page 1: ...Installation User Manual System manual E1 ...
Page 3: ...Copyright notice 4 System manual E1 This page intentionally left blank ...
Page 6: ...System manual E1 7 Preface ...
Page 14: ...System manual E1 15 Introduction ...
Page 20: ...System manual E1 21 Display controller installation ...
Page 26: ...System manual E1 27 Display installation ...
Page 34: ...System manual E1 35 Display controller software installation ...
Page 50: ...System manual E1 51 Display controller tools ...
Page 67: ...BarcoMed Hardware Tab This page intentionally left blank 68 System manual E1 ...
Page 68: ...System manual E1 69 NioWatch ...
Page 84: ...System manual E1 85 Display operation ...
Page 93: ...How to use the on screen display OSD This page intentionally left blank 94 System manual E1 ...
Page 94: ...System manual E1 95 Cleaning instructions ...
Page 99: ...LCD panel This page intentionally left blank 100 System manual E1 ...
Page 100: ...System manual E1 101 Troubleshooting ...
Page 112: ...System manual E1 113 Appendices ...
Page 123: ...Appendix C Warranty Statement 124 System manual E1 ...