7.3 Biological hazard and returns
Overview
The structure and the specifications of this device as well as the materials used for manufacturing makes it
easy to wipe and clean and therefore suitable to be used for various applications in hospitals and other
medical environments, where procedures for frequent cleaning are specified.
However, normal use shall exclude biological contaminated environments, to prevent spreading of infections.
Therefore use of this device in such environments is at the exclusive risk of Customer. In case this device is
used where potential biological contamination cannot be excluded.
Customer shall implement the decontamination process as defined in the latest edition of the ANSI/AAMI
ST35 standard on each single failed Product that is returned for servicing, repair, reworking or failure
investigation to Seller (or to the Authorized Service Provider). At least one adhesive yellow label shall be
attached on the top site of the package of returned Product and accompanied by a declaration statement
proving the Product has been successfully decontaminated.
Returned Products that are not provided with such external decontamination label, and/or whenever such
declaration is missing, can be rejected by Seller (or by the Authorized Service Provider) and shipped back at
Customer expenses.
7.4 Cleaning and disinfection
Instructions
•
Be sure to unplug the power cord from the mains when cleaning your LCD monitor.
•
Take care not to scratch the front surface with any hard or abrasive material.
•
Dust, finger marks, grease etc. can be removed with a soft damp cloth (a small amount of mild detergent
can be used on the damp cloth).
•
Wipe off water drop immediately.
Possible cleaning solutions
•
250 ppm chlorine solution
•
NaCI solution 0.9%
–
Sodim chloride 00-236
•
Bacillol AF
•
1.6 percent aqueous ammonia
•
Cidex® (2.4 percent glutaraldehyde solution)
•
Sodium hypochlorite (bleach) 10 percent
•
“
Green soap
”
(USP)
•
Like Cleansafe® optical cleaning liquid
•
Isopropanol
•
Haemosol solution (1% in 1 liter water)
•
Chlorehexidine 0,5% in 70% Ethanol
7.5 Regulatory compliance information
Indications for use
This device is intended to be used in operation rooms, to display images from endoscopic cameras, room and
boom cameras, ultrasound, cardiology, PACS, anesthesiology and patient information. It is not intended for
diagnosis.
Intended usage environment
•
Equipment primarily for use in a health care facility that is intended for use where contact with a patient is
unlikely (no applied part).
•
The equipment shall not be used with life support equipment.
Summary of Contents for MDSC-8231
Page 1: ...ENABLING BRIGHT OUTCOMES MDSC 8231 12G User Guide 31 UHD 4K surgical display MDSC 8231 ...
Page 5: ...R5911692 01 MDSC 8231 5 7 8 Legal disclaimer 59 7 9 Technical specifications 60 ...
Page 6: ...R5911692 01 MDSC 8231 6 ...
Page 7: ...7 R5911692 01 MDSC 8231 Welcome 1 ...
Page 10: ...R5911692 01 MDSC 8231 10 Welcome ...
Page 11: ...11 R5911692 01 MDSC 8231 Parts controls and connectors 2 ...
Page 17: ...17 R5911692 01 MDSC 8231 Installation 3 ...
Page 23: ...23 R5911692 01 MDSC 8231 Daily operation 4 ...
Page 28: ...R5911692 01 MDSC 8231 28 Daily operation ...
Page 43: ...43 R5911692 01 MDSC 8231 Troubleshooting 6 ...
Page 45: ...45 R5911692 01 MDSC 8231 Important information 7 ...