Bante810 Benchtop Dissolved Oxygen Meter
8
DO Calibration in % Saturation Mode
The Bante810 dissolved oxygen meter can perform 1 or 2 points calibration. For single point calibration,
we recommend that you perform 100% saturation
calibration in the air-saturated water. If the 2 points calibration is selected, the zero oxygen solution needs to be used.
Single point calibration - 100% saturation
1.1
Make sure that you have selected 1 point calibration in the setup menu.
1.2
Press the
Cal
key, the meter
shows “CAL1/100%”.
1.3
Hold the dissolved oxygen probe in the air at 100% relative humidity or place the probe into the air-saturated water for 15 minutes.
1.4
Press the
Enter
key. Wait for the reading to stabilize, the meter automatically shows END. Single point calibration is completed.
2 points calibration
2.1
Make sure that you have selected 2 points calibration in the setup menu.
2.2
Repeat the steps 1.2 to 1.4 above. When the first calibration point is completed, the display will
show “CAL2/0%”. The
meter prompts you to
continue with second point calibration.