Bante810 Benchtop Dissolved Oxygen Meter
9
2.3
Immerse the dissolved oxygen probe into the zero oxygen solution, stir the probe gently.
2.4
Press the
Enter
key to confirm. Wait for the reading to stabilize (this can often take more than 10 minutes), the display automatically shows END.
The meter returns to the measurement mode. Calibration is completed.
Performing a percentage saturation calibration will simultaneously calibrate the corresponding mg/L (or ppm) concentration value. Therefore,
additional mg/L calibration is not required in most circumstances.
Temperature Calibration
During the measurement process, if the temperature reading displayed differs from that of an accurate thermometer, the meter needs to be calibrated.
1.
Place the dissolved oxygen probe into a solution with a known accurate temperature.
2.
Press the
o
C
key to enter the temperature setting mode.
3.
Press the
or
key to set the temperature value.
4.
Press the
Enter
key, the meter returns to the measurement mode. Calibrating is completed.