Bante810 Benchtop Dissolved Oxygen Meter 

 

Installing the Electrode Holder 

Take out the electrode arm from the packaging. The base plate of the electrode holder has a circular hole, the electrode arm has a connecting rod. Insert the 
connecting rod into the circular hole and swivel the electrode arm 90

o

. Electrode holder is now ready to swing into desired position. 

 

   

 

 

Adjustment of electrode arm 

After installation, if the electrode arm automatically rises or falls, you need to adjust the screws until arm locate at any position.   
1. 

Remove the plastic cover from the electrode arm. 

2. 

Use the screwdriver to tighten the screw moderately. 

3. 

Insert the plastic cover to previous position. Installation is completed. 

 

   

 

 

Connecting the Dissolved Oxygen Probe 

1. 

Follow the steps below to place the probe into left or right side of the electrode arm. 

 
 

 

 
 

1. Insert the probe 

2. Hook up the cable 

3. Hook up the cable 

Summary of Contents for Bante810

Page 1: ...Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Page 2: ...rent work environment meets following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following li...

Page 3: ...ndicates the temperature compensation is enabled Setup mode icon Indicates the meter is in the setting mode Keypad The meter has a succinct membrane keypad names and symbols describe the each function...

Page 4: ...Solution 1 Take out the dissolved oxygen probe and electrolyte solution from the packaging Unscrew the membrane cap 2 Fill the membrane cap halfway with electrolyte solution 3 Screw the membrane cap o...

Page 5: ...o desired position Adjustment of electrode arm After installation if the electrode arm automatically rises or falls you need to adjust the screws until arm locate at any position 1 Remove the plastic...

Page 6: ...at its voltage matches the local main voltage 2 Insert the connector to the power socket The meter is now ready for use Switching the Meter On and Off Press the Meas key to switch on the meter the dis...

Page 7: ...meter will automatically sense a stable reading and lock the measurements Enable Disable Disable When the option is enabled the meter will automatically switch off if no key is pressed within 180 min...

Page 8: ...need to set a compatible parameter according to the local altitude ALTITUDE m kPa mmHg ALTITUDE m kPa mmHg 0 101 3 760 1600 82 9 622 100 100 1 750 1700 81 9 614 200 98 8 741 1800 80 9 607 300 97 6 73...

Page 9: ...1 point calibration in the setup menu 1 2 Press the Cal key the meter shows CAL1 100 1 3 Hold the dissolved oxygen probe in the air at 100 relative humidity or place the probe into the air saturated w...

Page 10: ...imultaneously calibrate the corresponding mg L or ppm concentration value Therefore additional mg L calibration is not required in most circumstances Temperature Calibration During the measurement pro...

Page 11: ...e 2 If necessary to set the barometric pressure and salinity coefficient in the setup menu Refer to page 7 3 Immerse the probe in the sample solution make sure the temperature sensor on the probe is f...

Page 12: ...Resolution 0 1mg L Saturation of Oxygen Range 0 0 200 0 Accuracy 2 0 Resolution 0 1 General Temperature Compensation 0 40 C 32 104 F Automatic Barometric Pressure Correction 60 0 112 5kPa 450 850mmHg...

Page 13: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...