Bante810 Benchtop Dissolved Oxygen Meter 

10 

 

Dissolved Oxygen Measurement 

The Bante810 dissolved oxygen meter is suitable for measuring the water, wastewater, brine and other liquids. If the sample is belong to the seawater or 
other water  containing large  amounts  of  salt,  please  setting  the  salinity  coefficient  before  measurement. Some gas  and  steam  such  as  chloride,  sulfur 
dioxide,  sulfureted  hydrogen,  ammonium,  carbon  dioxide  and  iodin  can  permeate  the  membrane  via  diffusion.  So  their  existence  will  influence  the 
measurement of dissolved oxygen. If the sample contains the solvent, grease, sulfide and alga, the membrane on the probe will be blocked, damaged or 
eroded. 
1. 

Connect the dissolved oxygen probe to meter and wait for 15 minutes to polarize the probe. 

2. 

If necessary, to set the barometric pressure and salinity coefficient in the setup menu (Refer to page 7). 

3. 

Immerse the probe in the sample solution, make sure the temperature sensor on the probe is fully immersed. 

4. 

Stir the probe gently. Record the measured value when the reading is stable. 

 

 

 

Auto-Hold 

The  meter  contains  an  Auto-Hold function.  If  enabled,  the  meter  will  automatically  sense  a  stable reading  and  lock  the measurements,  the  HOLD  icon 
appears on the display. If disabled, press the    key, the meter will immediately lock the displayed value. Press the 

Meas

 key to resume measuring. 

 

 

 

Electrode Care and Maintenance 

 

Always keep the membrane of the dissolved oxygen probe is wet or moist.   

 

If you do not use the probe for long periods,

 

please screw off membrane cap and rinse the cathode,

 

anode and membrane with deionized water, then 

soak up residual water on them with filter pape. Install the probe again. 

 

   

 

Temperature   
Sensor

 

Membrane cap 

Cathode and Anode 

Summary of Contents for Bante810

Page 1: ...Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Page 2: ...rent work environment meets following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following li...

Page 3: ...ndicates the temperature compensation is enabled Setup mode icon Indicates the meter is in the setting mode Keypad The meter has a succinct membrane keypad names and symbols describe the each function...

Page 4: ...Solution 1 Take out the dissolved oxygen probe and electrolyte solution from the packaging Unscrew the membrane cap 2 Fill the membrane cap halfway with electrolyte solution 3 Screw the membrane cap o...

Page 5: ...o desired position Adjustment of electrode arm After installation if the electrode arm automatically rises or falls you need to adjust the screws until arm locate at any position 1 Remove the plastic...

Page 6: ...at its voltage matches the local main voltage 2 Insert the connector to the power socket The meter is now ready for use Switching the Meter On and Off Press the Meas key to switch on the meter the dis...

Page 7: ...meter will automatically sense a stable reading and lock the measurements Enable Disable Disable When the option is enabled the meter will automatically switch off if no key is pressed within 180 min...

Page 8: ...need to set a compatible parameter according to the local altitude ALTITUDE m kPa mmHg ALTITUDE m kPa mmHg 0 101 3 760 1600 82 9 622 100 100 1 750 1700 81 9 614 200 98 8 741 1800 80 9 607 300 97 6 73...

Page 9: ...1 point calibration in the setup menu 1 2 Press the Cal key the meter shows CAL1 100 1 3 Hold the dissolved oxygen probe in the air at 100 relative humidity or place the probe into the air saturated w...

Page 10: ...imultaneously calibrate the corresponding mg L or ppm concentration value Therefore additional mg L calibration is not required in most circumstances Temperature Calibration During the measurement pro...

Page 11: ...e 2 If necessary to set the barometric pressure and salinity coefficient in the setup menu Refer to page 7 3 Immerse the probe in the sample solution make sure the temperature sensor on the probe is f...

Page 12: ...Resolution 0 1mg L Saturation of Oxygen Range 0 0 200 0 Accuracy 2 0 Resolution 0 1 General Temperature Compensation 0 40 C 32 104 F Automatic Barometric Pressure Correction 60 0 112 5kPa 450 850mmHg...

Page 13: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...

Reviews: