Bante810 Benchtop Dissolved Oxygen Meter 

 

2. 

Insert the 6-pin connector into the connector socket labeled DO. After the connection is completed, DO NOT pull on the cable. Always make sure that 
the connector is clean and dry. 

 

 

 
Connecting the Power Adapter 

1. 

Before plugging in the power adapter, ensure that its voltage matches the local main voltage. 

2. 

Insert the connector to the power socket. The meter is now ready for use. 

 

 

 

 

Switching the Meter On and Off 

 

Press the 

Meas

 key to switch on the meter, the display shows the measured values. 

 

Press and hold the 

Meas

 key for 5 seconds, the meter will switch off. 

 

  To enable the Auto-Power Off feature, please refer to chapter SETUP MENU. 

 

Prior to Use 

Connect the dissolved oxygen probe to meter. Turn on the meter 10 to 15 minutes and wait for the probe to polarize. 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for Bante810

Page 1: ...Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Page 2: ...rent work environment meets following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following li...

Page 3: ...ndicates the temperature compensation is enabled Setup mode icon Indicates the meter is in the setting mode Keypad The meter has a succinct membrane keypad names and symbols describe the each function...

Page 4: ...Solution 1 Take out the dissolved oxygen probe and electrolyte solution from the packaging Unscrew the membrane cap 2 Fill the membrane cap halfway with electrolyte solution 3 Screw the membrane cap o...

Page 5: ...o desired position Adjustment of electrode arm After installation if the electrode arm automatically rises or falls you need to adjust the screws until arm locate at any position 1 Remove the plastic...

Page 6: ...at its voltage matches the local main voltage 2 Insert the connector to the power socket The meter is now ready for use Switching the Meter On and Off Press the Meas key to switch on the meter the dis...

Page 7: ...meter will automatically sense a stable reading and lock the measurements Enable Disable Disable When the option is enabled the meter will automatically switch off if no key is pressed within 180 min...

Page 8: ...need to set a compatible parameter according to the local altitude ALTITUDE m kPa mmHg ALTITUDE m kPa mmHg 0 101 3 760 1600 82 9 622 100 100 1 750 1700 81 9 614 200 98 8 741 1800 80 9 607 300 97 6 73...

Page 9: ...1 point calibration in the setup menu 1 2 Press the Cal key the meter shows CAL1 100 1 3 Hold the dissolved oxygen probe in the air at 100 relative humidity or place the probe into the air saturated w...

Page 10: ...imultaneously calibrate the corresponding mg L or ppm concentration value Therefore additional mg L calibration is not required in most circumstances Temperature Calibration During the measurement pro...

Page 11: ...e 2 If necessary to set the barometric pressure and salinity coefficient in the setup menu Refer to page 7 3 Immerse the probe in the sample solution make sure the temperature sensor on the probe is f...

Page 12: ...Resolution 0 1mg L Saturation of Oxygen Range 0 0 200 0 Accuracy 2 0 Resolution 0 1 General Temperature Compensation 0 40 C 32 104 F Automatic Barometric Pressure Correction 60 0 112 5kPa 450 850mmHg...

Page 13: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...

Reviews: