Bante810 Benchtop Dissolved Oxygen Meter
6
Setup Menu
The Bante810 dissolved oxygen meter contains an integrated setup menu that is used to customize the displayed option to meet measurement requirements.
The following table describes the functions of the menu items.
MENU
DESCRIPTION
OPTIONS
DESCRIPTION
DEFAULT
Set the salinity coefficient of sample.
Range: 0~35ppt
0.0ppt
Set the barometric pressure coefficient.
Range: 450~850mmHg
760mmHg
Set the number of calibration points.
1 point
1 point
2 points
Set the default measurement unit.
mg/L
Concentration Unit
mg/L, ° C
ppm
%
% saturation
Degrees Celsius
Degrees Fahrenheit
When the option is enabled, the meter will automatically
sense a stable reading and lock the measurements.
Enable
Disable
Disable
When the option is enabled, the meter will automatically
switch off if no key is pressed within 180 minutes.
Enable
Disable
Disable
When the option is enabled, all of the calibration data and
selected parameters will back to factory default settings,
the meter must be recalibrated.
Enable
Disable
Disable
Setting a default option
1.
In the measurement mode, press and hold the
key for 3 seconds to enter the setup menu.
2.
Press the
or
key select the desired option.
3.
Press the
Enter
key to confirm and move to the next menu item.
4.
Repeat the steps above until the meter returns to the measurement mode. Setting is completed.
During the setting process, press the
Meas
key, the meter will exit the setting and return to the measurement mode.
Option
Menu
item