Bante810 Benchtop Dissolved Oxygen Meter 

 

Setting the Salinity Coefficient 

Salt dissolved in water will influences oxygen content of water. If your sample contains high levels of salinity, ensure that setting an applicable salinity   
coefficient prior to measurement. 
1.1

 

In the measurement mode, press and hold the

 

 

key for 3 seconds to enter the setup menu, the display shows

 

(salinity coefficient) option. 

1.2

 

Press the 

 or 

 key to set the value. 

1.3

 

Press the 

Enter

 key to confirm, the display shows the next menu item. 

1.4

 

Press the 

Meas

 key to exit the setting and return to the measurement mode. 

 

Setting the Barometric Pressure 

The following table describes the relationship between altitude and barometric pressure. Prior to the calibration or measurement, you need to set a 
compatible parameter according to the local altitude. 

 

ALTITUDE (m) 

kPa 

mmHg 

ALTITUDE (m) 

kPa 

mmHg 

101.3 

760 

1600 

82.9 

622 

100 

100.1 

750 

1700 

81.9 

614 

200 

98.8 

741 

1800 

80.9 

607 

300 

97.6 

732 

1900 

79.9 

599 

400 

96.4 

723 

2000 

78.9 

592 

500 

95.2 

714 

2100 

77.9 

584 

600 

94.0 

705 

2200 

76.9 

577 

700 

92.8 

696 

2300 

76.0 

570 

800 

91.7 

688 

2400 

75.0 

563 

900 

90.5 

679 

2500 

74.1 

556 

1000 

89.4 

671 

2600 

73.2 

549 

1100 

88.3 

662 

2700 

72.3 

542 

1200 

87.2 

654 

2800 

71.4 

536 

1300 

86.1 

646 

2900 

70.5 

529 

1400 

85.0 

638 

3000 

69.6 

522 

1500 

84.0 

630 

3100 

68.7 

515 

 

2.1 

In the measurement mode, press and hold the

 

 

key for 3 seconds to enter the setup menu, the display shows

 

(Salinity Coefficient) option. 

2.2 

Press the 

Enter

 key, the display shows

 

(Barometric Pressure Coefficient) option. 

2.3 

Press the 

 or 

 key to set the value. 

2.4 

Press the 

Enter

 key to confirm, the display shows the next menu item. 

2.5 

Press the 

Meas

 key to exit the setting and return to the measurement mode. 

 

 

 

 

 

 

Summary of Contents for Bante810

Page 1: ...Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Page 2: ...rent work environment meets following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following li...

Page 3: ...ndicates the temperature compensation is enabled Setup mode icon Indicates the meter is in the setting mode Keypad The meter has a succinct membrane keypad names and symbols describe the each function...

Page 4: ...Solution 1 Take out the dissolved oxygen probe and electrolyte solution from the packaging Unscrew the membrane cap 2 Fill the membrane cap halfway with electrolyte solution 3 Screw the membrane cap o...

Page 5: ...o desired position Adjustment of electrode arm After installation if the electrode arm automatically rises or falls you need to adjust the screws until arm locate at any position 1 Remove the plastic...

Page 6: ...at its voltage matches the local main voltage 2 Insert the connector to the power socket The meter is now ready for use Switching the Meter On and Off Press the Meas key to switch on the meter the dis...

Page 7: ...meter will automatically sense a stable reading and lock the measurements Enable Disable Disable When the option is enabled the meter will automatically switch off if no key is pressed within 180 min...

Page 8: ...need to set a compatible parameter according to the local altitude ALTITUDE m kPa mmHg ALTITUDE m kPa mmHg 0 101 3 760 1600 82 9 622 100 100 1 750 1700 81 9 614 200 98 8 741 1800 80 9 607 300 97 6 73...

Page 9: ...1 point calibration in the setup menu 1 2 Press the Cal key the meter shows CAL1 100 1 3 Hold the dissolved oxygen probe in the air at 100 relative humidity or place the probe into the air saturated w...

Page 10: ...imultaneously calibrate the corresponding mg L or ppm concentration value Therefore additional mg L calibration is not required in most circumstances Temperature Calibration During the measurement pro...

Page 11: ...e 2 If necessary to set the barometric pressure and salinity coefficient in the setup menu Refer to page 7 3 Immerse the probe in the sample solution make sure the temperature sensor on the probe is f...

Page 12: ...Resolution 0 1mg L Saturation of Oxygen Range 0 0 200 0 Accuracy 2 0 Resolution 0 1 General Temperature Compensation 0 40 C 32 104 F Automatic Barometric Pressure Correction 60 0 112 5kPa 450 850mmHg...

Page 13: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...