Bante810 Benchtop Dissolved Oxygen Meter
7
Setting the Salinity Coefficient
Salt dissolved in water will influences oxygen content of water. If your sample contains high levels of salinity, ensure that setting an applicable salinity
coefficient prior to measurement.
1.1
In the measurement mode, press and hold the
key for 3 seconds to enter the setup menu, the display shows
(salinity coefficient) option.
1.2
Press the
or
key to set the value.
1.3
Press the
Enter
key to confirm, the display shows the next menu item.
1.4
Press the
Meas
key to exit the setting and return to the measurement mode.
Setting the Barometric Pressure
The following table describes the relationship between altitude and barometric pressure. Prior to the calibration or measurement, you need to set a
compatible parameter according to the local altitude.
ALTITUDE (m)
kPa
mmHg
ALTITUDE (m)
kPa
mmHg
0
101.3
760
1600
82.9
622
100
100.1
750
1700
81.9
614
200
98.8
741
1800
80.9
607
300
97.6
732
1900
79.9
599
400
96.4
723
2000
78.9
592
500
95.2
714
2100
77.9
584
600
94.0
705
2200
76.9
577
700
92.8
696
2300
76.0
570
800
91.7
688
2400
75.0
563
900
90.5
679
2500
74.1
556
1000
89.4
671
2600
73.2
549
1100
88.3
662
2700
72.3
542
1200
87.2
654
2800
71.4
536
1300
86.1
646
2900
70.5
529
1400
85.0
638
3000
69.6
522
1500
84.0
630
3100
68.7
515
2.1
In the measurement mode, press and hold the
key for 3 seconds to enter the setup menu, the display shows
(Salinity Coefficient) option.
2.2
Press the
Enter
key, the display shows
(Barometric Pressure Coefficient) option.
2.3
Press the
or
key to set the value.
2.4
Press the
Enter
key to confirm, the display shows the next menu item.
2.5
Press the
Meas
key to exit the setting and return to the measurement mode.