Turbidity measuring

 TURBICUBE

 

BAMO IER GmbH 

 Pirnaer Straße 24 

 68309 Mannheim 

Tel. +49 (0)621 84224-0 

 Fax: +49 (0)621 84224-90 

. www.bamo.de 

 [email protected] 

SU3332a.doc  01/13 

3

 

The sensor set consists of a receiver module with microprocessor-controlled analysis electronics and two 
transmitter modules. 
The sensor set generates a 4 to 20 mA output signal in accordance with the following diagram. 
Measuring ranges are selected with the help of the DIP switches on the receiver module. 
 

Settings:

 

 

TURBICUBE 20 

TURBICUBE 1000 

DIP Switch Settings

 

Measuring 

Range

 

[FNU] 

DIP1  DIP2  DIP3  DIP4  DIP5  DIP6 *) 

1 = 0,01 ...    1  

ON 

ON  OFF  OFF  OFF 

OFF 

2 = 0,01 ...    2  

OFF  OFF  ON  OFF  OFF 

OFF 

3 = 0,01 ...    5  

ON  OFF  ON  OFF  OFF 

OFF 

4 = 0,01 ...   10   OFF  ON 

ON  OFF  OFF 

OFF 

5 = 0,01 ...   20  

ON 

ON 

ON  OFF  OFF 

OFF 

*) DIP 6 = ON 

 Colour-/ Fouling compensation is 

switched on  

DIP Switch Settings

 

Measuring  

Range

 

[FNU] 

DIP1  DIP2 

DIP3 

DIP4  DIP5*)  DIP6*) 

1 = 0,1 ...    50 

OFF  OFF  OFF 

ON 

OFF 

OFF 

2 = 0,1 ...  100 

ON  OFF  OFF 

ON 

OFF 

OFF 

3 = 0,1 ...  200 

OFF  ON 

OFF 

ON 

OFF 

OFF 

4 = 0,1 ...  500 (**300)  ON 

ON 

OFF 

ON 

OFF 

OFF 

5 = 100 ...1000 

OFF  OFF  OFF  OFF 

OFF 

OFF 

*)   DIP-switch 5 and 6 without function,  
Normal position = OFF 

**) armatures with nominal diameter > DN65 - DN100

 

 

 

Troubleshooting: 

 

Measuring current 22 mA  = overflow 

 

Measuring current 0 mA = wrong DIP-switch setting (no valid measuring range selected) 

 

Indication 

green LED = 1 Hz blinking 

ready, measuring in operation 

green LED = permanent ON 

ready, measuring not in operation 

green LED = OFF 

no power supply or defective 

Fault indication

 

red LED = ON 

= measuring circuit faulty, (short circuit or broken) 

red LED = 1 Hz blinking 
only TURBICUBE 20 

= glasses dirty or liquids to much colorized  
(damping factor > 20 dB) 

 

Note:  

The turbidity measuring instruments are calibrated with the internationally specified standard suspension 

Formazin

. The indication takes place 

thus not in form of the measured light intensity, but as concentration of the 

Formazin

 calibration suspension. During the measurement of any 

liquid this means thus that the liquid concerned causes the same light scattering as the standard suspension of the indicated concentration. 
Measured values of other turbidity measuring instruments, which use other calibration suspensions and measuring angle, cannot be compared 
directly with those with Formazin calibrated measuring instruments! 
 

Summary of Contents for IER Turbicube 1000

Page 1: ...Functions Description The turbidity measuring device determines the turbidity of liquids using the light absorption and scattered light measuring method The inline measuring cell at the sensor set mak...

Page 2: ...mark if the measuring circuit is also connected to the auxiliary power supply of the load will increase to additional 24 VDC x 20mA 0 48W Terminal housing PBT IP 65 Ambient temperature 5 45 C Medium t...

Page 3: ...OFF OFF OFF OFF OFF DIP switch 5 and 6 without function Normal position OFF armatures with nominal diameter DN65 DN100 Troubleshooting Measuring current 22 mA overflow Measuring current 0 mA wrong DI...

Page 4: ...Module PCB Elektrical Connection Observe The measuring output 4 20 mA is a current drain that means the output channel has the same electrical behavior as a variable resistor A active power supply fo...

Page 5: ...e fixture by unscrewing the sleeve nuts screw the ETR glasses with the included tool and remove clean the lens do not use abrasive cleansers which may cause scratching tighten the ETR glasses with the...

Page 6: ...ng is required as part of the respective quality assurance system for test equipment monitoring calibration can be checked with the calibration rods and the device can be readjusted if necessary A sui...

Page 7: ...bration rod from the box Insert the calibration rod into the fixture Mount the S2 transmitter to the calibration rod observe the locking pin Insert the calibration rod with attached transmitter comple...

Page 8: ...SW1 setpoint on the plate in the box of the KN D calibration rod Actual value measured value Setpoint actual value deviation Less than 5 measuring instrument is OK Greater than 5 adjust actual value...

Page 9: ...onnection both side flange DIN 2501 DN20 d25 240 240 300 300 DN25 d32 240 240 300 300 DN32 d40 240 240 300 300 DN40 d50 240 240 300 300 DN50 d63 262 262 342 300 DN65 d75 314 314 342 300 DN80 d90 347 3...

Reviews: