PERATING 

I

NSTRUCTIONS

Turbidity measuring

 TURBICUBE

 

BAMO IER GmbH 

 Pirnaer Straße 24 

 68309 Mannheim 

Tel. +49 (0)621 84224-0 

 Fax: +49 (0)621 84224-90 

. www.bamo.de 

 [email protected] 

SU3332a.doc  01/13 

9

 

 

Mounting distance: 

 

  

PVC 

 

  (tolerance: –0/+2 mm- ) 

PP-type 

(tolerance: ±2 mm ) 

nominal dia (DN) 

process conn. 

d both 

side 

thread 

connection 

both side flange 

DIN 2501 

both side thread 

connection 

both side flange 

DIN 2501 

DN20 

d25 240 

240 

300 

300 

DN25 

d32 240 

240 

300 

300 

DN32 

d40 240 

240 

300 

300 

DN40 

d50 240 

240 

300 

300 

DN50 

d63 262 

262 

342 

300 

DN65 

d75 314 

314 

342 

300 

DN80 

d90 347 

347 

394 

400 

DN100 

d110 347 

347 

394 

400 

 

Observe!         Only for vertical ascending tubes! 
Mounting in a ascending tube with slow down section in front and behind the 
TURBICUBE is optimal. 

 

Dimensions 

 

Example: 
TURBICUBE FF ..2

 ( PP-type with flanges on both sides , flanges acc. ISO 7005, EN 1092, DIN 2501 PN10 

 

Example: 
TURBICUBE VV .. 1   

(PVC-type with DN-threading on both sides)

 

Summary of Contents for IER Turbicube 1000

Page 1: ...Functions Description The turbidity measuring device determines the turbidity of liquids using the light absorption and scattered light measuring method The inline measuring cell at the sensor set mak...

Page 2: ...mark if the measuring circuit is also connected to the auxiliary power supply of the load will increase to additional 24 VDC x 20mA 0 48W Terminal housing PBT IP 65 Ambient temperature 5 45 C Medium t...

Page 3: ...OFF OFF OFF OFF OFF DIP switch 5 and 6 without function Normal position OFF armatures with nominal diameter DN65 DN100 Troubleshooting Measuring current 22 mA overflow Measuring current 0 mA wrong DI...

Page 4: ...Module PCB Elektrical Connection Observe The measuring output 4 20 mA is a current drain that means the output channel has the same electrical behavior as a variable resistor A active power supply fo...

Page 5: ...e fixture by unscrewing the sleeve nuts screw the ETR glasses with the included tool and remove clean the lens do not use abrasive cleansers which may cause scratching tighten the ETR glasses with the...

Page 6: ...ng is required as part of the respective quality assurance system for test equipment monitoring calibration can be checked with the calibration rods and the device can be readjusted if necessary A sui...

Page 7: ...bration rod from the box Insert the calibration rod into the fixture Mount the S2 transmitter to the calibration rod observe the locking pin Insert the calibration rod with attached transmitter comple...

Page 8: ...SW1 setpoint on the plate in the box of the KN D calibration rod Actual value measured value Setpoint actual value deviation Less than 5 measuring instrument is OK Greater than 5 adjust actual value...

Page 9: ...onnection both side flange DIN 2501 DN20 d25 240 240 300 300 DN25 d32 240 240 300 300 DN32 d40 240 240 300 300 DN40 d50 240 240 300 300 DN50 d63 262 262 342 300 DN65 d75 314 314 342 300 DN80 d90 347 3...