Procedure for TURBICUBE 1000: 

 
The following adjustment sequence must be adhered to!! 
 
Step 1 

 

Clean all lenses and wipe them dry (all visible water droplets and water film must be removed!). 

 

Then reinstall the lenses for the receiver and the S1 transmitter 

only

 

Do not yet install

 the lens and the compression disc for the S2 transmitter (180° scattered light)!! 

 

Reinstall the S1 transmitter and the receiver. 

 

Remove the KN-D calibration rod (through light) 

from the box. 

 

Insert the calibration rod into the fixture 

 

Mount the S2 transmitter to the calibration rod (observe the locking pin) 

 

 

 

 

Insert the calibration rod with attached transmitter completely into the fixture (observe the locking pin),  
and tighten the sleeve nut. 

 

Set the DIP switches to the MB5 range = 100 to 1000 (all DIP-switches OFF). 

 

Switch supply power back on again – the status LED blinks! 

 

Compare the setpoint value and the actual value. 
Setpoint 1 

(SW1)

 = setpoint on the plate in the box of the 

 KN-D

 calibration rod 

Actual value = measured value 

 

Setpoint / actual value deviation: 
Less than ±5% 

 measuring instrument is OK 

Greater than ±5% 

 adjust actual value with potentiometer P3 

 Remove 

the

 KN-D calibration rod (through light)

Step 2 

 

Remove the KN-S calibration rod (scattered light)

 from the box. 

 

Mount the calibration rod to the S2 transmitter as described above. 

 

DIP switches stay on the MB5 range = 100 to 1000 (all switches off). 

 

Compare the setpoint value and the actual value. 
Setpoint 2 

(SW2)

 = setpoint on the plate in the box of the 

 KN-S

 calibration rod 

Actual value = measured value  

 

Setpoint / actual value deviation: 
Less than ±5% 

 measuring instrument is OK 

Greater than ±5% 

 adjust actual value with potentiometer P2 

Step 3 

 

Set the DIP switches to the MB4 range = 0,1 to 500 (DIP-switches ON/ON/OFF/ON/OFF/OFF). 

 

Compare the setpoint value and the actual value. 
Setpoint 3 

(SW3)

 = setpoint on the plate in the box of the

 KN-D

 calibration rod 

Actual value = measured value 

 

Setpoint / actual value deviation: 
Less than ±5% 

 measuring instrument is OK 

Greater than ±5% 

 adjust actual value with potentiometer P4 

 Remove 

the

 KN-D calibration rod

Step 4 

 

mount glass with sealing ring and compression disc and transmitter S2 

 

select the right DIP-switch setting for operation 

Summary of Contents for IER Turbicube 1000

Page 1: ...Functions Description The turbidity measuring device determines the turbidity of liquids using the light absorption and scattered light measuring method The inline measuring cell at the sensor set mak...

Page 2: ...mark if the measuring circuit is also connected to the auxiliary power supply of the load will increase to additional 24 VDC x 20mA 0 48W Terminal housing PBT IP 65 Ambient temperature 5 45 C Medium t...

Page 3: ...OFF OFF OFF OFF OFF DIP switch 5 and 6 without function Normal position OFF armatures with nominal diameter DN65 DN100 Troubleshooting Measuring current 22 mA overflow Measuring current 0 mA wrong DI...

Page 4: ...Module PCB Elektrical Connection Observe The measuring output 4 20 mA is a current drain that means the output channel has the same electrical behavior as a variable resistor A active power supply fo...

Page 5: ...e fixture by unscrewing the sleeve nuts screw the ETR glasses with the included tool and remove clean the lens do not use abrasive cleansers which may cause scratching tighten the ETR glasses with the...

Page 6: ...ng is required as part of the respective quality assurance system for test equipment monitoring calibration can be checked with the calibration rods and the device can be readjusted if necessary A sui...

Page 7: ...bration rod from the box Insert the calibration rod into the fixture Mount the S2 transmitter to the calibration rod observe the locking pin Insert the calibration rod with attached transmitter comple...

Page 8: ...SW1 setpoint on the plate in the box of the KN D calibration rod Actual value measured value Setpoint actual value deviation Less than 5 measuring instrument is OK Greater than 5 adjust actual value...

Page 9: ...onnection both side flange DIN 2501 DN20 d25 240 240 300 300 DN25 d32 240 240 300 300 DN32 d40 240 240 300 300 DN40 d50 240 240 300 300 DN50 d63 262 262 342 300 DN65 d75 314 314 342 300 DN80 d90 347 3...

Reviews: