8
Procedure for TURBICUBE 1000:
The following adjustment sequence must be adhered to!!
Step 1
Clean all lenses and wipe them dry (all visible water droplets and water film must be removed!).
Then reinstall the lenses for the receiver and the S1 transmitter
only
.
Do not yet install
the lens and the compression disc for the S2 transmitter (180° scattered light)!!
Reinstall the S1 transmitter and the receiver.
Remove the KN-D calibration rod (through light)
from the box.
Insert the calibration rod into the fixture
Mount the S2 transmitter to the calibration rod (observe the locking pin)
Insert the calibration rod with attached transmitter completely into the fixture (observe the locking pin),
and tighten the sleeve nut.
Set the DIP switches to the MB5 range = 100 to 1000 (all DIP-switches OFF).
Switch supply power back on again – the status LED blinks!
Compare the setpoint value and the actual value.
Setpoint 1
(SW1)
= setpoint on the plate in the box of the
KN-D
calibration rod
Actual value = measured value
Setpoint / actual value deviation:
Less than ±5%
measuring instrument is OK
Greater than ±5%
adjust actual value with potentiometer P3
Remove
the
KN-D calibration rod (through light)
.
Step 2
Remove the KN-S calibration rod (scattered light)
from the box.
Mount the calibration rod to the S2 transmitter as described above.
DIP switches stay on the MB5 range = 100 to 1000 (all switches off).
Compare the setpoint value and the actual value.
Setpoint 2
(SW2)
= setpoint on the plate in the box of the
KN-S
calibration rod
Actual value = measured value
Setpoint / actual value deviation:
Less than ±5%
measuring instrument is OK
Greater than ±5%
adjust actual value with potentiometer P2
Step 3
Set the DIP switches to the MB4 range = 0,1 to 500 (DIP-switches ON/ON/OFF/ON/OFF/OFF).
Compare the setpoint value and the actual value.
Setpoint 3
(SW3)
= setpoint on the plate in the box of the
KN-D
calibration rod
Actual value = measured value
Setpoint / actual value deviation:
Less than ±5%
measuring instrument is OK
Greater than ±5%
adjust actual value with potentiometer P4
Remove
the
KN-D calibration rod
.
Step 4
mount glass with sealing ring and compression disc and transmitter S2
select the right DIP-switch setting for operation