background image

 

O

PERATING 

I

NSTRUCTIONS

Turbidity measuring

 TURBICUBE

 

BAMO IER GmbH 

 Pirnaer Straße 24 

 68309 Mannheim 

Tel. +49 (0)621 84224-0 

 Fax: +49 (0)621 84224-90 

. www.bamo.de 

 [email protected] 

SU3332a.doc  01/13 

5

 

 

Maintenance: 

Cleaning intervals depend upon the type of liquid medium and must be determined by the user. 
 
Cleaning the lenses: 

 

Disconnect the turbidity-measuring instrument from all sources of power. 

 

Depressurize pipes and empty liquid medium from the device. 

 

Remove the transmitters and the receiver from the fixture by unscrewing the sleeve nuts. 

 

 

 

screw the ETR-glasses with the 
included tool and remove. 

clean the lens – 

do not use abrasive cleansers, 

which may cause scratching!!

 

 

 

  

 

 

tighten the ETR-glasses with the included tool . 

 

Before reinstalling the transmitters and the receiver, refill the fixture with liquid medium and inspect for 
possible leaks. 

 

 

 

   Observe positioning of hole and pin!

 

 

Observe for cold fluids 

The fixture has to be installed to the pipelines 

with mounted sensors

 only 

For cleaning the glasses 

dismount the complete fixture

, after cleaning install as described above 

 

Note:  
Inspect the O-ring before reassembly 
and replace if necessary. 
 
mount the o-ring approximately as 
shown in the picture 

 

Summary of Contents for IER Turbicube 1000

Page 1: ...Functions Description The turbidity measuring device determines the turbidity of liquids using the light absorption and scattered light measuring method The inline measuring cell at the sensor set mak...

Page 2: ...mark if the measuring circuit is also connected to the auxiliary power supply of the load will increase to additional 24 VDC x 20mA 0 48W Terminal housing PBT IP 65 Ambient temperature 5 45 C Medium t...

Page 3: ...OFF OFF OFF OFF OFF DIP switch 5 and 6 without function Normal position OFF armatures with nominal diameter DN65 DN100 Troubleshooting Measuring current 22 mA overflow Measuring current 0 mA wrong DI...

Page 4: ...Module PCB Elektrical Connection Observe The measuring output 4 20 mA is a current drain that means the output channel has the same electrical behavior as a variable resistor A active power supply fo...

Page 5: ...e fixture by unscrewing the sleeve nuts screw the ETR glasses with the included tool and remove clean the lens do not use abrasive cleansers which may cause scratching tighten the ETR glasses with the...

Page 6: ...ng is required as part of the respective quality assurance system for test equipment monitoring calibration can be checked with the calibration rods and the device can be readjusted if necessary A sui...

Page 7: ...bration rod from the box Insert the calibration rod into the fixture Mount the S2 transmitter to the calibration rod observe the locking pin Insert the calibration rod with attached transmitter comple...

Page 8: ...SW1 setpoint on the plate in the box of the KN D calibration rod Actual value measured value Setpoint actual value deviation Less than 5 measuring instrument is OK Greater than 5 adjust actual value...

Page 9: ...onnection both side flange DIN 2501 DN20 d25 240 240 300 300 DN25 d32 240 240 300 300 DN32 d40 240 240 300 300 DN40 d50 240 240 300 300 DN50 d63 262 262 342 300 DN65 d75 314 314 342 300 DN80 d90 347 3...

Reviews: