Maintenance:
Cleaning intervals depend upon the type of liquid medium and must be determined by the user.
Cleaning the lenses:
Disconnect the turbidity-measuring instrument from all sources of power.
Depressurize pipes and empty liquid medium from the device.
Remove the transmitters and the receiver from the fixture by unscrewing the sleeve nuts.
screw the ETR-glasses with the
included tool and remove.
clean the lens –
do not use abrasive cleansers,
which may cause scratching!!
tighten the ETR-glasses with the included tool .
Before reinstalling the transmitters and the receiver, refill the fixture with liquid medium and inspect for
possible leaks.
Observe positioning of hole and pin!
Observe for cold fluids
The fixture has to be installed to the pipelines
with mounted sensors
only
For cleaning the glasses
dismount the complete fixture
, after cleaning install as described above
Note:
Inspect the O-ring before reassembly
and replace if necessary.
mount the o-ring approximately as
shown in the picture