Testing and Readjustment
Factory Calibration
The sensor set is equipped with a calibration constant as a design feature.
As a rule, no recalibration is required.
Calibration point accuracy is better than 2%‚ and drift is less than 1% per year.
Test Equipment Monitoring
If device calibration testing is required as part of the respective quality assurance system for test equipment
monitoring, calibration can be checked with the calibration rods, and the device can be readjusted if necessary.
A suitable calibration standard is delivered with each turbidity measuring instrument.
Each calibration standard is matched to the specific circumstances of the mating turbidity measuring instrument,
and cannot be used for other measuring instruments of the same type!
The calibration standard and the turbidity measuring instrument must have the same serial number!
Calibration box
with mounting tool and calibration rod for TURBICUBE 20
(for TURBICUBE 1000 there are two rods in the box)