BG 800 300 BE / A (0105) TPG 300
30
4.7.9 »Parameter storage« group
The stored parameters are activated when the TPG 300 is
switched on. If no parameters have been stored, the unit defaults
to the standard parameter set given in section 4.7.3.
a) »Parameter set« function
1 2 3 4 A B
0
10
You can either select your own set of parameters (user) or the
default set to be saved.
Display
Meaning
SA u
SA H
SA d
Save user parameters (
SA
ve
u
ser)
Save user parameters with immediate start-up
(
SA
ve
H
otstart)
Save default (factory set) parameters(
SA
ve
d
e-
faults)
By activating the immediate start-up, a measuring circuit
can be automatically re-enabled after a power failure. This
is particularly useful in the case of self-monitoring.
The immediate start-up is jointly activated for all measur-
ing circuits. The measuring circuit must however be
switched on during storage.
At «SA u» and «SA H», the unit can be locked by a code
number (
→
following section).
b) »Save« function
Saving the default parameters has the following ef-
fects:
•
The switching function assignments are lost.
•
The relays are de-energized, i.e. the switching
functions change to »off«.
•
Communication with a computer may no longer be
possible.
1 2 3 4 A B
0
10