BG 800 300 BE / A (0105) TPG 300
41
6.2.4 Replacing the apparatus fuses
The fuses are located inside the unit on the power supply board.
Procedure
•
Switch the unit off and wait one minute
•
Remove all cables (mains last)
•
Unscrew the back panel
•
Pull out the back panel together with the plug-in boards
just far enough to make the fuses accessible
•
Replace the fuses
Make sure to replace fuses with fuses of the recom-
mended type and nominal current strength (
→
2
62). It
is not permissible to use mended fuses or to short circuit
the fuse holder.
•
Reassemble in reverse order
Put the screw lockings in place again and
tighten the screws!
If the fuses blow again.