BG 800 300 BE / A (0105) TPG 300
27
c) »Threshold setting« function
1 2 3 4 A B
mbar
0
10
<step>
<funct>
<funct>
Modifications only become effective when the switching
function, group or operating mode is changed.
d) »Measuring circuit assignment« function
Any of the switching points can be assigned to any of the
measuring channels.
Changing the assignment can trigger a change in the
switching function status.
1 2 3 4 A B
mbar
A1
A2
B1
B2
0
10
The cycle depends on the plug-in boards installed.
Available measuring circuits are indicated by an LED.
The upper and lower thresholds of switching functions
1 ... 4 cannot be assigned to different measuring circuits.
The last entry made applies.
The upper and lower thresholds of switching functions A
and B can be assigned to different measuring circuits
(
→
section 4.5.4).
The LED for the assigned measuring circuit flashes.
It is possible to leave a switching function unassigned (no
measuring circuit LED will flash). The switching function is
ineffective.
Modifications only become effective when the switching
function, group or operating mode is changed.