BG 800 300 BE / A (0105) TPG 300
36
6 Error messages, troubleshooting
Manipulations inside the unit that are not described in this
operating manual may only be performed by a Balzers
Service Center.
6.1 Error messages
An error message is
indicated by a light-
ing or flashing «Err»
LED.
1 2 3 4 A B
mbar
A2
Err
0
10
Display
Possible cause
Correction
«dt»
3)
Watch Dog timer – overflow due to strong external influ-
ence (electromagnetic)
<funct>
If this error occurs frequently, replace the basic unit
«EE»
3)
Error during parameter reading
Store default or user parameters(
→
section 4.7.9)
EEPROM defective
«EP»
3)
EPROM defective
«Id»
3)
Operating system overloaded
<funct>
«IF»
3)
Interface plug-in board in slot A or B
Put the interface plug-in board into slot C
2)