BIS U-602_ Processor Unit
TCP/IP / RS232
36
Reading 8 bytes of user data (1122334455667788) from TAG with
EPC 0102030405060708090A0B0C
hex
from address 0.
1. Command
: 'RE'000000000C0B0A090807060504030201'
˽
U000000000008
˽
˽
'<LF><CR>
'RE' 00…01 '
˽
U 000000 000008
˽
˽
' <LF><CR>
Reading on data carrier
with EPC
EPC of data carrier
1)
No TID, only user data
Start address
No. of bytes
Reserved
End identifier
2. Reply:
<ACK>002C'RE'000000000C0B0A090807060504030201'
˽
U000000000008'
00001122334444667788<LF><CR>
<ACK> 002C 'RE' 00…01 '
˽
U 00…08' 0...0 11…88 <LF><CR>
Acknowledgment
Reply length of 44 bytes
Reading on data carrier
with EPC
EPC of data carrier
1)
No TID, only user data
Start number of
bytes
Reserved (0000
hex
)
Data block
2)
End identifier
1) Fully recognized EPC: 0102030405060708090A0B0C
hex
, filled to 16 bytes with 00000000
hex
.
2) Data block 8 bytes: Address 0: 11
hex
, Address 1: 22
hex,
Address 3: 33
hex
,... Address 7: 88
hex
.
Telegram
example
Read 'R'
8
Device Function
Summary of Contents for BIS U-6020-053-10 RS232