Sash Opening
(Inches)
Model
Intake airflow at 80 fpm [0.406 m/sec]
(cfm)
8 [203.2mm]
SG403A
205 [96.8 L/sec]
SG603A
311 [147 L/sec]
10 [254mm]
SG403A
256 [121 L/sec]
SG603A
389 [184 L/sec]
12 [304.8mm]
SG403A
306 [144 L/sec]
SG603A
467 [220 L/sec]
Cabinet intake airflow values at 80 fpm [0.406 m/sec]
1)
Low Flow Alarm Set Point: The ADJUST screw on the AFM box wants to be set so the yellow light
will come on, then 6 seconds later the red light will turn on activating the audible alarm. (Make sure
the slide switch for the audible alarm is in the ON position). Turning the alarm ADJUST screw
clockwise will move the CFM alarm point up and counter clockwise will move the CFM alarm point
down.
2)
Return the exhaust system to it proper flow rate and recheck cabinet intake velocity for HEC
operation.
3)
Return the cabinet blower motor speed controller to it original voltage and recheck cabinet intake
velocity for CEC operation.
NOTE: If an exhaust transition or cabinet damper adjustment is performed during a routine cabinet
certification, the AFM alarm should be recalibrated using the steps above.
(For either the Exhaust to Room, CEC or HEC connection).
Summary of Contents for SterilGARD III Advance
Page 35: ...Document 333D012G 5 18 2007 34 Appendix...
Page 38: ...Document 333D012G 5 18 2007 37 Dimensional Drawing Standard Configuration SG403A...
Page 39: ...Document 333D012G 5 18 2007 38 Dimensional Drawing Standard Configuration SG603A...
Page 41: ...Document 333D012G 5 18 2007 40 Interior View Exterior View Cable port plug not shown...
Page 42: ...Document 333D012G 5 18 2007 41 Ladder Schematic SG403A...
Page 43: ...Document 333D012G 5 18 2007 42 Ladder Schematic SG603A...