Document: 333D012G
5-18-2007
- 49 -
4)
Remove
FLUOR-A-LOCK
tabs (2) from the lamp sockets located on the left side of the fluorescent
lamps. (Fig. 3) Using both hands, grasp the lamp carefully near each end and rotate it approx ¼ turn
until the lamp is no longer seated in the lamp sockets. Slide each end out of the sockets. Repeat the
process for removing the second lamp and dispose of them properly.
5)
Install new lamps by following step 4 in reverse order. Make sure that the locking disks are installed on
one end of each lamp and that the
FLUOR-A-LOCK
tabs are inserted properly. (Fig. 3)
6)
Hinge light canopy up temporarily and press the light on/off switch to verify that both lamps are
working.
7)
Lower canopy halfway and re-insert restraint cables on either end of electrical mounting board. Close
the canopy and install the thumbscrews finger tight.
8)
While lifting the vertical dress panel up slightly, hinge the dress panel standoff up against the sloped
dress panel. Gently lower the vertical dress panel down until it rests against the cabinet.
Ultraviolet Lamp Replacement
As reported in other sections of this manual, UV germicidal lamps lose their effectiveness over time and
should be replaced when the intensity at the work surface drops below 40 microwatts per square centimeter
at a wavelength of 253.5 x 10-9 meters.
Fig. 3
Summary of Contents for SterilGARD III Advance
Page 35: ...Document 333D012G 5 18 2007 34 Appendix...
Page 38: ...Document 333D012G 5 18 2007 37 Dimensional Drawing Standard Configuration SG403A...
Page 39: ...Document 333D012G 5 18 2007 38 Dimensional Drawing Standard Configuration SG603A...
Page 41: ...Document 333D012G 5 18 2007 40 Interior View Exterior View Cable port plug not shown...
Page 42: ...Document 333D012G 5 18 2007 41 Ladder Schematic SG403A...
Page 43: ...Document 333D012G 5 18 2007 42 Ladder Schematic SG603A...