Document: 333D012G
5-18-2007
- 48 -
Replacing Fluorescent and Ultraviolet Lamps
Fluorescent Lamp Replacement
1)
Turn the fluorescent lamps off by depressing the
light on/off switch locating on the right side of
the light canopy.
2)
Pull the vertical panel out enough at the bottom
so that the thumbscrews can be accessed. (Fig.1)
Hinge the dress panel standoff (located on the
left side of the sloped dress panel) down and
lower the vertical panel on it. While removing
the thumbscrews, support the light canopy with
the other hand. Lower the canopy until the
restraint cables hold it in place.
3)
Disconnect the restraint cables from the face of
the electrical mounting board at both ends of the
light canopy. (Fig. 2) Gently hinge the canopy
down until it rests on the front of the unit.
V er tica l
P a n el
T h u m b
S cr ew
F ig . 1
L ig h t
C a n op y
Fig. 2
Light
Restraint