Document: 333D012G
5-18-2007
- 53 -
EXHAUSTING REQUIREMENTS FOR HEC AND CEC
Cabinet
Model
Transition
Model
Sash
Height
(Inches)
*Exhaust
Flow Rate
(cfm)
Pressure
Loss
(” w.c.)
Added Height to Type A
Cabinet with ET
ATD / SHORT COLLAR
(Inchs)
Exhaust
Duct
Diameter
(Inchs)
SG403A HEC403A
8
[203.2mm]
269
[127 L/sec]
0.06
[14.9 Pa]
11-1/2 / 5-5/16
[292 / 135mm]
10
[254mm]
SG403A HEC403A
10
[254mm]
335
[158 L/sec]
0.08
[19.9 Pa]
11-1/2 / 5-5/16
[292 / 135mm]
10
[254mm]
SG603A HEC603A
8
[203.2mm]
408
[193 L/sec]
0.10
[24.9 Pa]
11-1/2 / 5-5/16
[292 / 135mm]
10
[254mm]
SG603A HEC603A
10
[254mm]
510
[241 L/sec]
0.15
[37.4 Pa]
11-1/2 / 5-5/16
[292 / 135mm]
10
[254mm]
Exhaust requirements for Hard Exhaust Connection (HEC)
* = All exhaust flow rates are based on a nominal 105 fpm [0.533 m/sec] intake velocity.