Document: 333D012G
5-18-2007
- 45 -
Only the Adjustable Leg Leveler is shown for clarity
5)
If your unit was purchased with the
Caster Option
be sure to
chock the previously
installed
casters
(
to prevent cabinet movement
) then repeat steps 3 and 4 for the other end of the unit.
Raising the Unit to 30 1/8” [765mm] Work Surface Height - Adjustable Leg Leveler Option
Raising the Unit to 30” [762mm] Work Surface Height - Caster Option
6)
Lift one end of the unit up a minimum of 5” [127mm] and block in position.
7)
Unscrew the bolts (One per leg) that hold the legs in the shipping position.
8)
Slide the leg out of the stand until the holes line up at the next position.
2,3,4