Chapter 3. Operation and Programming
50
Moisture Target Series 6 User’s Manual
3.6.6 Reading and Setting the Calibration References
Note:
The following procedure applies only if an
M Series
probe is being used. If a
VeriDri
probe is used,
Cal Reference...
is not accessible.
IMPORTANT:
The
MTS6
is factory programmed with high and low reference MH values. These values are generated
from a factory lab calibration and should not be changed without first consulting Panametrics technical
support. Changes to these values will alter the accuracy of the unit measurements.
To update calibration data, from the
Settings Menu
select
Cal Data
and press
Enter
.
The following screen appears.
Note:
To access the
Cal Data
menu, the
User Passcode
is required (see section 3.6.1 on
page 43).
Use the left and right arrow keys to select each digit to be changed and the up and
down arrow keys to increase or decrease its value. Once the passcode has been
entered, press
Enter
a screen similar to the following appears.
To view and/or edit the calibration reference settings, select
Cal Reference
and press
Enter
. The following screen appears.
Settings Menu
Cal Data...
AutoCal...
Fault Alm...
V/V Ratio...
Clock...
Probes
DpC Offset
0000
/=Save =Cancel
User Passcode:
Summary of Contents for Panametrics MTS6
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 23: ...Moisture Target Series 6 User s Manual 15 Chapter 2 Installation Figure 27 MTS6 Wiring Diagram...
Page 32: ...Chapter 2 Installation 24 Moisture Target Series 6 User s Manual...
Page 73: ...Moisture Target Series 6 User s Manual 65 Chapter 4 Service and Maintenance...
Page 78: ...Chapter 5 Specifications 70 Moisture Target Series 6 User s Manual...
Page 83: ...Chapter B Menu Maps Moisture Target Series 6 Hygrometer User s Manual93 Chapter B Menu Maps...
Page 98: ...76 Moisture Target Series 6 User s Manual...
Page 104: ...Warranty 2 Moisture Target Series 6 User s Manual no content intended for this page...
Page 105: ......