Moisture Target Series 6 User’s Manual
31
Chapter 3. Operation and Programming
3.3.5 Changing the Lower Output Span
To adjust the lower output span, from the
Output Menu
select
Lower
and press
Enter
. A screen similar to the following appears.
Use the left and right arrow keys to select each digit to be changed and the up and
down arrow keys to increase or decrease its value. Press
Enter
to save (or
Cancel
to keep the previous value), and return to
Output Menu
.
3.3.6 Testing the Output
The
Test Menu
causes the
MTS6
to generate a 0- or 4-20mA output, or a 0-2V output, at the percent of scale selected.
For example, in 4-20mA operation, 0% = 4mA, 50% = 12mA, 100% = 20mA. This allows the proper function of recording
or SCADA equipment to be verified. In 0-20 operation, 0% = 0mA, 50% = 10mA, 100% = 20mA
To test system output, from the
Output Menu
select
Test
and press
Enter
. The
MTS6
will
proceed to check the settings, and a screen similar to the following will appear.
Use the left and right arrow keys to select each digit to be changed, and the up and
down arrow keys to increase or decrease its value. Press
Enter
to save (or
Cancel
to keep the previous value), and return to the
Output Menu
.
Check your output wiring. If the reading on your SCADA or DCS is off slightly, then you
may use the
Trim
feature to trim the output zero or span.
3.3.7 Trimming the Outputs
The
Trim Menu
enables the operator to compensate for differences in measurement of the 0/4-20 mA or 0-2V outputs
by connected recorders or SCADA equipment. To trim the output:
Select
Trim
from the
Output Menu
and press
Enter
. The following screen appears.
When performing a
Trim
operation, the
MTS6
unit requires you to first reset the trim. To reset the trim output, select
Reset Trim
and press
Enter
. The following screen appears.
Type
Units
Upper
Test
Trim...
Output Menu
Lower
–110.0 DP °C
/=Save =Cancel
+60.0
Set Output Zero
Max:
–110.0
Min:
Type
Units
Upper
Lower
Trim...
Output Menu
Test
+050.00 %
/=Apply =Exit
+110.00
Output Test Value:
Max:
–25.00
Min:
Type
Units
Upper
Test
Upper
Output Menu
Trim...
Output Trim
Reset Trim
Trim Zero
Trim Span
Summary of Contents for Panametrics MTS6
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 23: ...Moisture Target Series 6 User s Manual 15 Chapter 2 Installation Figure 27 MTS6 Wiring Diagram...
Page 32: ...Chapter 2 Installation 24 Moisture Target Series 6 User s Manual...
Page 73: ...Moisture Target Series 6 User s Manual 65 Chapter 4 Service and Maintenance...
Page 78: ...Chapter 5 Specifications 70 Moisture Target Series 6 User s Manual...
Page 83: ...Chapter B Menu Maps Moisture Target Series 6 Hygrometer User s Manual93 Chapter B Menu Maps...
Page 98: ...76 Moisture Target Series 6 User s Manual...
Page 104: ...Warranty 2 Moisture Target Series 6 User s Manual no content intended for this page...
Page 105: ......