Chapter 2. Installation
4
Moisture Target Series 6 User’s Manual
2.2 Selecting the Recorder Output
Note:
By default, the recorder is set to the current output.
Note:
The customer will provide their own cable for connecting the recorder. Acceptable cables range from 16 to
26AWG.
The
MTS6
has one isolated analog recorder output. The recorder output provides either a current or voltage signal,
which is set by switch
S1
on the main PC board.
Complete these steps to check or reset switch
S1
(see Figure 8).
WARNING!
Never connect line voltage or any other power input to the recorder output terminals.
1.
Make sure the
MTS6
is turned off and unplugged.
WARNING!
The MTS must be isolated or disconnected from all voltage sources before changing the recorder
output.
2.
Remove the screw at the top of the back panel (see Figure 4).
Figure 4: Back Panel
3.
Lift the back edge of the cover (see Figure 5), slide the cover back (see Figure 6), and lift it from the enclosure
(see Figure 7).
Figure 5: Lifting the Back Edge of the Cover
Summary of Contents for Panametrics MTS6
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 23: ...Moisture Target Series 6 User s Manual 15 Chapter 2 Installation Figure 27 MTS6 Wiring Diagram...
Page 32: ...Chapter 2 Installation 24 Moisture Target Series 6 User s Manual...
Page 73: ...Moisture Target Series 6 User s Manual 65 Chapter 4 Service and Maintenance...
Page 78: ...Chapter 5 Specifications 70 Moisture Target Series 6 User s Manual...
Page 83: ...Chapter B Menu Maps Moisture Target Series 6 Hygrometer User s Manual93 Chapter B Menu Maps...
Page 98: ...76 Moisture Target Series 6 User s Manual...
Page 104: ...Warranty 2 Moisture Target Series 6 User s Manual no content intended for this page...
Page 105: ......