Chapter 3. Operation and Programming
32
Moisture Target Series 6 User’s Manual
Use the left or right arrow keys to select
YES
and press
Enter
. This cancels any
previous trim values, and returns the
MTS6
to its factory adjustment. The display
returns to the previous screen.
To trim the zero value, select
Trim Zero
and press
Enter
. A screen similar to the
following appears.
This will cause the
MTS6
to output 4.000 mA or 0.4 V on the output being trimmed. The
output value should then be read using the connected recorder, SCADA equipment, or
DVM. Enter the value read from the connected equipment as the
Zero Trim
value, as follows:
Note:
Since you cannot trim 0 mA or 0 V for negative offsets, trim for the lower end of the scale is at the 4 mA or
0.4 V output level.
Use the left and right arrow keys to select each digit to be changed, and the up and
down arrow keys to increase or decrease its value. Press
Enter
to save (or
Cancel
to keep the previous value).
The Output
Trim
menu returns with
Trim Span
highlighted. To change the span value,
press
Enter
. A screen similar to the following appears.
This will cause the
MTS6
to output 20.000 mA on the output being trimmed. The output
value should then be read using the connected recorder, SCADA equipment, or DVM.
Enter the value read from the connected equipment as the
Span Trim
value.
Use the left and right arrow keys to select each digit to be changed, and the up and
down arrow keys to increase or decrease its value. Press
Enter
to save (or
Cancel
to keep the previous value).
Trimming is complete. Accuracy can be verified using the
Test Menu
, above.
Example:
Trim
is reset, then
Trim Zero
is selected. The SCADA input reports 3.977 mA.
The operator enters “3.977” as the
Zero Trim
value.
Trim Span
is selected. The SCADA input reports 19.985 mA.
The operator enters “19.985” as the
Span Trim
value.
MTS6
will adjust the output accordingly to true the output as read by the customer recorder, SCADA or DVM.
Using the
Test Menu
, the operator verifies that a test value of 0% now reads 4.000 mA at the SCADA equipment, and a
test value of 100% now reads 20.000 mA.
3.4 Setting Up Alarms
Note:
The
MTS6
is equipped with two programmable high/low alarm relays and one fault alarm.
3.4.1
Selecting an Alarm Output
To set up alarm outputs, on the
Main Menu
choose
Alarm
and press
Enter
. From the
Alarm Menu
choose
Select
and press
Enter
. A screen similar to the following appears.
Summary of Contents for Panametrics MTS6
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 23: ...Moisture Target Series 6 User s Manual 15 Chapter 2 Installation Figure 27 MTS6 Wiring Diagram...
Page 32: ...Chapter 2 Installation 24 Moisture Target Series 6 User s Manual...
Page 73: ...Moisture Target Series 6 User s Manual 65 Chapter 4 Service and Maintenance...
Page 78: ...Chapter 5 Specifications 70 Moisture Target Series 6 User s Manual...
Page 83: ...Chapter B Menu Maps Moisture Target Series 6 Hygrometer User s Manual93 Chapter B Menu Maps...
Page 98: ...76 Moisture Target Series 6 User s Manual...
Page 104: ...Warranty 2 Moisture Target Series 6 User s Manual no content intended for this page...
Page 105: ......