62 |
© 2021 Baker Hughes Company. All rights reserved.
Table 9: Input Signal and Power, Specifications
Table 10: Construction Material Specifications
Table 11: System Connectivity
Power Supply
Loop powered from 4-20 mA control signal
Compliance Voltage Rating
9.0 V at 20 mA, 11.0 V at 4.0 mA
Minimum Current Signal to Start Up
3.2 mA
Minimum Input Span for Split
Range Operation
5 mA
Upper Range Value for Split
Range Operation
8 mA to 20 mA
Lower Range Value for Split
Range Operation
4 mA to 14 mA
Wire Size
12/28 AWG
Strip Length
0.43 in / 11 mm
Digital Communication
HART
®
Communication protocol revision
5 (firmware version 2.2.1) and 7 (firmware
version 3.1.1 and later).
Housing and Cover
Low Copper Aluminum Alloy
Weight
o
SVi1000: 3.2 lbs./ 1.451 kg
o
SVi1000 SW/G/IM: 4.1 lbs./ 1.860 kg
Relay
Nitrile diaphragms, Polycarbonate
I/P Motor
430 stainless steel, Low Copper Aluminum Alloy, 300 series
stainless steel, nitrile diaphragm
Magnet Holder
Corrosion Protected Anodized Aluminum 6061 T6
Pole Ring
416 stainless steel
Levers
300 Series stainless steel
HART
®
Physical Device Type
Positioner; HART
®
cmd rev 5 or 7, Device type 204
(0x00cc)
DD Registered with the Field
Comm
®
Group
Yes
Integration with HART
®
Host
software
ValVue standalone, ValVue AMS SNAP-ON application
available, Plug-In Application For Yokogawa
®
PRM,
ValVue For Honeywell
®
FDM, Device Type Manager
(DTM) for FDT
®
Host