I/O (INPUT / OUTPUT) GROUP
Main
Adv
Cal
System
Setup
Meter
Log
I/O
Test Out
Set Out
Scale Out
Set In
Figure 77: I/O group
Set Out (Setup Outputs) Page
Figure 78: Data outputs setup
The
I/O > Set Out
option determines which of the flow meter’s outputs are active .
• Current Output Mode
• Voltage Output Mode
• Digital Output Mode
The DXN flow meter’s output signals are analog or digital/pulse . Analog signals change continuously over time . Digital/
pulse signals are present at discrete points and represent information using a sequence of on/off pulses . For connection
information, see
.
Select the type of I/O from the pull-down lists . The two analog outputs are 4…20 mA and 0…10V DC output . The pulse
output has a maximum frequency of 1000 Hz and can be configured as either a rate pulse or a totalizing pulse .
All outputs are scaled the same way using the
I/O > Scale Out
. For example, 20 mA, 1000 Hz and 0…10V all represent the same
maximum flow rate set in the
I/O > Scale Out.