H25-IR
Operation
Instruction 3015-5286
51
5.4.5.3 IR Emitter Adjustment
WARNING!
SHOCK HAZARD
.
When performing the following step, be careful not to come in contact
with AC voltages when working inside the instrument with the cover off and AC voltage applied.
1. Reconnect the AC power cord, and then turn the instrument ON while holding down the front panel
ENT
button to enable the Factory menu
2. Wait until the instrument warms up and the Data Display screen appears before proceeding.
3. Press
the
ENT
and
ESC
buttons simultaneously to access the Main menu. From the Main menu,
select DIAGNOSTICS > IR EMITTER to display the IR Emitter Parameters screen.
4. Adjust potentiometer R37 located on the main board for an IR Emitter reading of
800 ±20
or
450
±20 mW
, corresponding to the user’s version of part as seen in the items required section
containing the part numbers.
5.4.5.4 DIGIPOT Adjustment
1. Press
ESC
twice to return to the Main menu. From the Main menu, select FACTORY > DIGIPOT to
display the DIGIPOT screen.
IMPORTANT!
When performing the following Step, the instrument must be sampling fresh air
(zero gas) when adjusting the DIGIPOT value.
2. Use
the
Up
and
Down
buttons to adjust the DIGIPOT value until the IR VOLT reading is
4.2000 ±0.0500 volts. Press
ENT
to save the new DIGIPOT value.
IMPORTANT!
If adjusting the DIGIPOT does not reach 4.2000 ±0.0500 volts then return to step 18 to
lower the IR Emitter setting. If the IR Emitter setting is 450 ±20 mW adjust the IR Emitter
setting to 375 ±20 mW. Next, repeat the DIGIPOT adjustment.
3. Turn OFF instrument and remove AC power cord.
4. Reassemble
instrument.
IR EMITTER PARAMETERS
VOLTS=5.7345 mA=77.5
OHMS=75.0
mW=450.0
DIGIPOT=220
IR VOLT=4.2000
NOTE:
Readings for the other
parameters on this screen may vary.
Summary of Contents for H25-IR
Page 6: ...Contents H25 IR iv Instruction 3015 5286 Notes ...
Page 26: ...Preparing for Operation H25 IR 20 Instruction 3015 5286 Notes ...
Page 34: ...Calibration H25 IR 28 Instruction 3015 5286 Notes ...
Page 43: ...H25 IR Operation Instruction 3015 5286 37 Notes ...
Page 68: ...Specifications H25 IR 62 Instruction 3015 5286 Notes ...