Maintenance
H25-IR
38
Instruction 3015-5286
5 Maintenance
5.1 Diagnostics Menu
The Diagnostics menu provides information about the instrument’s operation that will be useful to service
technicians for the purpose of troubleshooting various fault conditions.
To display the Diagnostics menu, first press
both
the
ENT
and
ESC
buttons on either the instrument’s front
panel or Smart probe at the same time to display the Main menu. Next, use the
Up
and
Down
buttons to
highlight “DIAGNOSTICS” and then press
ENT
to display the Diagnostics menu.
Main Menu:
SYSTEM
LOG
SETUPS
CAL
DATA
DIAGNOSTICS
5.1.1 Sensor Data
The Sensor Data screen provides information about the sensor’s operating parameters and its surrounding
environment. When contacting the Bacharach Service Department, the operator may be asked to provide this
information to help in the troubleshooting process.
To display the Sensor Data screen, first display the Diagnostics menu as described in Section 5.1. Next, use
the
Up
and
Down
buttons to highlight “SENSOR DATA” and then press
ENT
.
Sensor Data Screen Information:
LR
..........................Measured Leak Rate
FAULT
...................Current Numerical Fault Code (refer to Section
5.1.2 Last Fault
)
OPEN/CLOSED
.....Measure/Search Flow Rate Valve Position
DET
........................Detector Voltage
N
............................Detector’s Output Noise Factor (with no gas detected)
Z
.............................Stored Zero Voltage of Detector
DET
........................Detector Temperature
SPPM
......................PPM Corrected to Standard Pressure and Temperature
BOX
.......................Enclosure Temperature
M
............................Manifold Pressure
AMB
........................Ambient Pressure
MPM
........................Moles Per Minute
AU
...........................Average Optical Absorbency Unit of Gas Being Detected (0 = no gas)
PPM
........................Parts Per Million Value of the Current Detected Gas Level
UML
........................Absolute Concentration of Gas in Micro Moles per Liter
Diagnostics Menu:
SENSOR DATA
LAST
FAULT
IR
EMITTER
PRESSURE
LOOP
TEST
CLR LAST FAULT
AUTO
GAIN
Summary of Contents for H25-IR
Page 6: ...Contents H25 IR iv Instruction 3015 5286 Notes ...
Page 26: ...Preparing for Operation H25 IR 20 Instruction 3015 5286 Notes ...
Page 34: ...Calibration H25 IR 28 Instruction 3015 5286 Notes ...
Page 43: ...H25 IR Operation Instruction 3015 5286 37 Notes ...
Page 68: ...Specifications H25 IR 62 Instruction 3015 5286 Notes ...