5.1.3 IR Emitter
The IR Emitter screen provides readouts of the detector’s voltage, current, resistance, and power consumption.
When contacting the Bacharach Service Department, the operator may be asked to provide this information to
help in the troubleshooting process.
To display the IR Emitter screen, first display the Diagnostics menu as described in Section 5.1. Next, use the
Up
and
Down
buttons to highlight “IR EMITTER” and then press
ENT
.
5.1.4 Pressure
The Pressure screen provides readouts of bench pressure, ambient pressure, the difference between the two
(delta), flow rate through the bench in moles per minute, and the tank pressure of the internal leak source.
When contacting the Bacharach Service Department, the operator may be asked to provide this information to
help in the troubleshooting process.
To display the Pressure screen, first display the Diagnostics menu as described in Section 5.1. Next, use the
Up
and
Down
buttons to highlight “PRESSURE” and then press
ENT
.
This screen also allows a technician to manually turn the pump ON and OFF, along with manually controlling
the position of the measure/search flow-rate valve.
Pressing the
Down
button toggles the pump ON and OFF.
Pressing the
Left
button opens and closes the measure/search flow-rate valve. An asterisk (
) on the display
indicates that the valve is in its measure position.
5.1.5 Loop Test
The Loop Test screen allows the operator to force the instrument’s 4–20 mA loop output to 0, 4, 10, or 20 mA.
This function is useful for checking the calibration of external monitoring equipment such as chart recorders.
To display the Loop Test screen, first display the Diagnostics menu as described in Section 5.1. Next, use the
Up
and
Down
buttons to highlight “LOOP TEST” and then press
ENT
.
Press the
Up
,
Down
,
Left
, and
Right
buttons to produce the desired output current. Press
ESC
to return to the
Diagnostics menu.
5.1.6 Clear Last Fault
The Last Fault screen normally displays the last fault detected by the instrument, even if the fault no longer
exists. This is useful in determining the cause of an intermittent type fault.
To clear the Last Fault screen of all faults: first display the Diagnostics menu as described in Section 5.1; use
the
Up
and
Down
buttons to highlight “CLR LAST FAULT”; press
ENT
; and then confirm this action by pressing
the
Up
button.
IR EMITTER PARAMETERS
VOLTS=6.1740 mA=129.5
OHMS=47.7 mW=800.0
BENCH
AMB DELT
11.73 14.04
2.31
FLOW=0.00950Mpm
LEAK PSI=102.1(100.2)
LOOP TEST
DOWN=0ma UP=20ma
LEFT=4ma RIGHT=10ma
ESC TO QUIT
Summary of Contents for H25-IR
Page 6: ...Contents H25 IR iv Instruction 3015 5286 Notes ...
Page 26: ...Preparing for Operation H25 IR 20 Instruction 3015 5286 Notes ...
Page 34: ...Calibration H25 IR 28 Instruction 3015 5286 Notes ...
Page 43: ...H25 IR Operation Instruction 3015 5286 37 Notes ...
Page 68: ...Specifications H25 IR 62 Instruction 3015 5286 Notes ...