Parts &Service
H25-IR
46
Instruction 3015-5286
10. Select
Setup > File Locations
and verify that the
coldload.bin, pilot.bin, and flash.ini files are all located in the
C:\RFU folder.
Click
OK
if these files are in the correct folder. If necessary,
use the browse buttons in the Choose File Locations dialog
box to locate each of these files.
11. Select
File > Load Flash Image
.
12. Click the browse button in the Choose Flash Image
dialog box; navigate to the C:\RFU folder; and then
highlight the bin file of the new firmware’s flash image
file. In this example, the file
H25IR 1_20.bin
will be
loaded.
Click
Open
.
13. Click
OK
to begin the flashing process.
14. A progress-dialog box will appear during the flashing
process, and will disappear when the flashing process is
complete.
15. When flashing is complete: Exit the flash program; turn
OFF the instrument; remove the serial cable; and reset
the DIP switches back to their normal operation
positions.